A formalism for independent checking of Gamma Knife dose calculations

被引:14
作者
Tsai, JS
Engler, MJ
Rivard, MJ
Mahajan, A
Borden, JA
Zheng, Z
机构
[1] Tufts New England Med Ctr, Boston Gamma Knife Ctr, Dept Radiat Oncol, Boston, MA 02111 USA
[2] Tufts New England Med Ctr, Boston Gamma Knife Ctr, Dept Neurosurg, Boston, MA 02111 USA
[3] Brown Univ, Sch Med, New England Gamma Knife Ctr, Dept Radiat Oncol, Providence, RI 02903 USA
关键词
Gamma Knife; radiosurgery; stereotactic radiosurgery; dose verification; quality assurance; off-center-ratio; output factor; isocenter; dosimetric verification;
D O I
10.1118/1.1398038
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
For stereotactic radiosurgery using the Leksell Gamma Knife system, it is important to perform a pre-treatment verification of the maximum dose calculated with the Leksell GammaPlan((R)) (D-LGP) stereotactic radiosurgery system. This verification can be incorporated as part of a routine quality assurance (QA) procedure to minimize the chance of a hazardous overdose. To implement this procedure, a formalism has been developed to calculate the dose D-CAL(X,Y,Z,d(av),t) using the following parameters: average target depth (d(av)), coordinates (X,Y,Z) of the maximum dose location or any other dose point(s) to be verified, 3-dimensional (3-dim) beam profiles or off-center-ratios (OCR) of the four helmets, helmet size i, output factor O-i, plug factor Pi, each shot j coordinates (x,y,z)(i,j), and shot treatment time (t(i.j)). The average depth of the target d(av) was obtained either from MRI/CT images or ruler measurements of the Gamma Knife Bubble Head Frame. D-CAL and D-LGP were then compared to evaluate the accuracy of this independent calculation. The proposed calculation for an independent check of D-LGP has been demonstrated to be accurate and reliable. and thus serves as a QA tool for Gamma Knife stereotactic radiosurgery. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:1842 / 1849
页数:8
相关论文
共 9 条
[1]
STEREOTACTIC HEAVY-ION BRAGG PEAK RADIOSURGERY FOR INTRA-CRANIAL VASCULAR DISORDERS - METHOD FOR TREATMENT OF DEEP ARTERIOVENOUS-MALFORMATIONS [J].
FABRIKANT, JI ;
LYMAN, JT ;
HOSOBUCHI, Y .
BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (678) :479-490
[2]
KJELLBERG RN, 1977, NEUROLOGICAL SURG, P181
[3]
HIGH-ENERGY PROTON BEAM AS A NEUROSURGICAL TOOL [J].
LARSSON, B ;
LEKSELL, L ;
REXED, B ;
SOURANDER, P ;
MAIR, W ;
ANDERSSON, B .
NATURE, 1958, 182 (4644) :1222-1223
[4]
STEREOTACTIC RADIOSURGERY [J].
LEKSELL, L .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1983, 46 (09) :797-803
[5]
LEKSELL L, 1951, ACTA CHIR SCAND, V102, P316
[6]
A SYSTEM FOR STEREOTACTIC RADIOSURGERY WITH A LINEAR-ACCELERATOR [J].
LUTZ, W ;
WINSTON, KR ;
MALEKI, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (02) :373-381
[7]
GammaPlan®-Leksell Gamma Knife® radiosurgery treatment planning verification method [J].
Marcu, SM ;
Wu, QJ ;
Pillai, K ;
Weinhous, MS .
MEDICAL PHYSICS, 2000, 27 (09) :2146-2149
[8]
THE SHEFFIELD STEREOTAXIC RADIOSURGERY UNIT - PHYSICAL CHARACTERISTICS AND PRINCIPLES OF OPERATION [J].
WALTON, L ;
BOMFORD, CK ;
RAMSDEN, D .
BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (717) :897-906
[9]
ZHANG P, 2001, J APPL CLIN MED PHYS, V1, P158