Detection of chemical weapon agents and simulants using chemical ionization reaction time-of-flight mass Spectrometry

被引:31
作者
Cordell, Rebecca L. [1 ]
Willis, Kerry A. [1 ]
Wyche, Kevin P. [1 ]
Blake, Robert S. [1 ]
Ellis, Andrew M. [1 ]
Monks, Paul S. [1 ]
机构
[1] Univ Leicester, Dept Chem, Leicester LE1 7RH, Leics, England
基金
英国自然环境研究理事会;
关键词
D O I
10.1021/ac071193c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Chemical ionization reaction time-of-flight mass spectrometry (CIR-TOF-MS) has been used for the analysis of prepared mixtures of chemical weapon agents (CWAs) sarin and sulfur mustard. Detection of the CWA simulants 2-chloroethyl ethyl sulfide, triethyl phosphate, and dimethyl methyl phosphonate has also been investigated. Chemical ionization of all the agents and simulants was shown to be possible using the CIR-TOF-MS technique with a variety of reagent ions, and the sensitivity was optimized by variation of instrument parameters. The ionization process was found to be largely unaffected by sample humidity levels, demonstrating the potential suitability of the method to a range of environmental conditions, including the analysis of CWAs in air and in the breath of exposed individuals.
引用
收藏
页码:8359 / 8366
页数:8
相关论文
共 40 条
[1]  
[Anonymous], CONV PROH DEV PROD S
[2]   APPLICATION OF GAS-CHROMATOGRAPHY MASS-SPECTROMETRY AND GAS-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRY TO THE ANALYSIS OF CHEMICAL WARFARE SAMPLES, FOUND TO CONTAIN RESIDUES OF THE NERVE AGENT SARIN, SULFUR MUSTARD AND THEIR DEGRADATION PRODUCTS [J].
BLACK, RM ;
CLARKE, RJ ;
READ, RW ;
REID, MTJ .
JOURNAL OF CHROMATOGRAPHY A, 1994, 662 (02) :301-321
[3]   Analysis of degradation products of organophosphorus chemical warfare agents and related compounds by liquid chromatography mass spectrometry using electrospray and atmospheric pressure chemical ionisation [J].
Black, RM ;
Read, RW .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :233-244
[4]   Chemical ionization reaction time-of-flight mass spectrometry: Multi-reagent analysis for determination of trace gas composition [J].
Blake, Robert S. ;
Wyche, Kevin P. ;
Ellis, Andrew M. ;
Monks, Paul S. .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2006, 254 (1-2) :85-93
[5]   Demonstration of proton-transfer reaction time-of-flight mass spectrometry for real-time analysis of trace volatile organic compounds [J].
Blake, RS ;
Whyte, C ;
Hughes, CO ;
Ellis, AM ;
Monks, PS .
ANALYTICAL CHEMISTRY, 2004, 76 (13) :3841-3845
[6]   Packed capillary liquid chromatography-electrospray ionization (tandem) mass spectrometry of mustard hydrolysis products in soil [J].
D'Agostino, PA ;
Hancock, JR ;
Chenier, CL .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1058 (1-2) :97-105
[7]   Mass spectrometric analysis of chemical warfare agents and their degradation products in soil and synthetic samples [J].
D'Agostino, PA ;
Hancock, JR ;
Chenier, CL .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2003, 9 (06) :609-618
[8]   Determination of sarin, soman and their hydrolysis products in soil by packed capillary liquid chromatography-electrospray mass spectrometry [J].
D'Agostino, PA ;
Hancock, JR ;
Provost, LR .
JOURNAL OF CHROMATOGRAPHY A, 2001, 912 (02) :291-299
[9]  
DWYER A, 2003, JANES CHEMBIO HDB
[10]   Ion mobility spectrometers in national defense. [J].
Eiceman, GA ;
Stone, JA .
ANALYTICAL CHEMISTRY, 2004, 76 (21) :390A-397A