Novel hexacyanoferrate (III)-modified carbon electrodes: Application in miniaturized biosensors with potential for in vivo glucose sensing

被引:44
作者
Jaffari, SA [1 ]
Pickup, JC [1 ]
机构
[1] GUYS HOSP, UMDS, DIV CHEM PATHOL, LONDON SE1 9RT, ENGLAND
关键词
miniature glucose biosensor; glucose oxidase; 4-acetamidophenol; acetaminophen; paracetamol; ascorbic acid; modified carbon; potassium hexacyanoferrate (III);
D O I
10.1016/0956-5663(96)82339-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We adapted a new technology for the modification of carbon base electrodes for use as probe-type, potentially-implantable glucose sensors. Carbon rods (diameter 0.3 mm) were modified by repeated potential cycling in 0.1 M potassium hexacyanoferrate (III). The modified-carbon electrodes were sealed in plastic pipette tips with an exposed reaction area where glucose oxidase was immobilized using glutaraldehyde. An outer membrane of Nafion, followed by 15% (w/v) polyurethane, was applied over the enzyme layer. The miniature modified-carbon glucose sensors displayed a sensitivity to glucose in phosphate-buffered saline of 91.4+/-19 nA/mM (mean+/-SEM) and a linear range up to 5.3+/-1 mM glucose when operated at 750 mV versus a silver/silver chloride reference. Corresponding, unmodified-carbon based glucose sensors displayed a lower sensitivity of 20.7+/-3 nA/mM with a linear range up to 3.8+/-0.5 mM. The modified-carbon glucose sensors responded to glucose when operated in plasma but with a reduced sensitivity compared with that in buffered saline. Glucose sensors displayed good stability for up to 6.5 days during continuous operation in 5 mM buffered glucose solution. Interference from ascorbate and 4-acetamidophenol at both physiological and pharmacological ranges was significantly lower at the modified-carbon base electrodes than that at the unmodified-carbon base electrodes. Also, the relatively large effect of ascorbate and 4-acetamidophenol at the unmodified-carbon base electrode was reduced considerably when the base electrode was coated with glucose oxidase, Nafion and polyurethane membranes. (C) 1996 Elsevier Science Limited
引用
收藏
页码:1167 / 1175
页数:9
相关论文
共 23 条
  • [1] DESIGN AND INVITRO STUDIES OF A NEEDLE-TYPE GLUCOSE SENSOR FOR SUBCUTANEOUS MONITORING
    BINDRA, DS
    ZHANG, YN
    WILSON, GS
    STERNBERG, R
    THEVENOT, DR
    MOATTI, D
    REACH, G
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (17) : 1692 - 1696
  • [2] FERROCENE-MEDIATED ENZYME ELECTRODE FOR AMPEROMETRIC DETERMINATION OF GLUCOSE
    CASS, AEG
    DAVIS, G
    FRANCIS, GD
    HILL, HAO
    ASTON, WJ
    HIGGINS, IJ
    PLOTKIN, EV
    SCOTT, LDL
    TURNER, APF
    [J]. ANALYTICAL CHEMISTRY, 1984, 56 (04) : 667 - 671
  • [3] CSOREGI E, 1994, ANAL CHEM, V66, P3131
  • [4] EVALUATION OF NAFION AS MEDIA FOR GLUCOSE-OXIDASE IMMOBILIZATION FOR THE DEVELOPMENT OF AN AMPEROMETRIC GLUCOSE BIOSENSOR
    FORTIER, G
    VAILLANCOURT, M
    BELANGER, D
    [J]. ELECTROANALYSIS, 1992, 4 (03) : 275 - 283
  • [5] IMMOBILIZATION OF GLUCOSE-OXIDASE IN FERROCENE-MODIFIED PYRROLE POLYMERS
    FOULDS, NC
    LOWE, CR
    [J]. ANALYTICAL CHEMISTRY, 1988, 60 (22) : 2473 - 2478
  • [6] TWO-DIMENSIONAL ENZYME ELECTRODE SENSOR FOR GLUCOSE
    GOUGH, DA
    LUCISANO, JY
    TSE, PHS
    [J]. ANALYTICAL CHEMISTRY, 1985, 57 (12) : 2351 - 2357
  • [7] CROSS-LINKED REDOX GELS CONTAINING GLUCOSE-OXIDASE FOR AMPEROMETRIC BIOSENSOR APPLICATIONS
    GREGG, BA
    HELLER, A
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (03) : 258 - 263
  • [8] ROLE OF CARBONYL SITES IN THE ELECTROCHEMICAL RESPONSE OF 2,6-DIAMINOPURIN-8-OL
    BODALBHAI, L
    BRAJTERTOTH, A
    [J]. ANALYTICA CHIMICA ACTA, 1991, 243 (01) : 33 - 42
  • [9] POLYFERROCENES AS MEDIATORS IN AMPEROMETRIC BIOSENSORS FOR GLUCOSE
    HENDRY, SP
    CARDOSI, MF
    TURNER, APF
    NEUSE, EW
    [J]. ANALYTICA CHIMICA ACTA, 1993, 281 (03) : 453 - 459
  • [10] IMMOBILIZED ENZYME CHEMICALLY MODIFIED ELECTRODE AS AN AMPEROMETRIC SENSOR
    IANNIELLO, RM
    YACYNYCH, AM
    [J]. ANALYTICAL CHEMISTRY, 1981, 53 (13) : 2090 - 2095