Simple method for modulation transfer function determination of digital imaging detectors from edge images

被引:47
作者
Buhr, E [1 ]
Günther-Kohfahl, S [1 ]
Neitzel, U [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
来源
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2003年 / 5030卷
关键词
modulation transfer function; edge image analysis; digital radiography; digital x-ray detector;
D O I
10.1117/12.479990
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A simple variant of the edge method to determine the presampled modulation transfer function (MTF) of digital imaging detectors has been developed that produces sufficiently accurate MTF values for frequencies up to the Nyquist frequency limit of the detector with only a small amount of effort for alignment and computing. An oversampled edge spread function (ESF) is generated from the image of a slanted edge by rearranging the pixel data of N consecutive lines that correspond to a lateral shift of-the edge of one pixel. The original data are used for the computational analysis without further data preprocessing. Since the number of lines leading to an edge shift of one pixel is generally a fractional number rather than an integer, a systematic error may. be introduced in the MTF obtained. Simulations and theoretical investigations show that for all frequencies up to the Nyquist limit the relative error DeltaMTF/MTF is below 1/(2N) and can thus be kept below a given threshold by a suitable choice of N. The method is especially useful for applications where the MTF is needed for frequencies up to the Nyquist frequency limit, like the determination of the detective quantum efficiency (DQE).
引用
收藏
页码:877 / 884
页数:8
相关论文
共 13 条
[1]   SIGNAL AND NOISE IN MODULATION TRANSFER-FUNCTION DETERMINATIONS USING THE SLIT, WIRE, AND EDGE TECHNIQUES [J].
CUNNINGHAM, IA ;
REID, BK .
MEDICAL PHYSICS, 1992, 19 (04) :1037-1044
[2]  
Dobbins J., 2000, HDB MED IMAGING, V1
[3]  
FISHER MG, 1982, PHOTOGR SCI ENG, V26, P32
[4]   A SIMPLE METHOD FOR DETERMINING THE MODULATION TRANSFER-FUNCTION IN DIGITAL RADIOGRAPHY [J].
FUJITA, H ;
TSAI, DY ;
ITOH, T ;
DOI, K ;
MORISHITA, J ;
UEDA, K ;
OHTSUKA, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) :34-39
[5]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .1. MODULATION TRANSFER-FUNCTION [J].
GIGER, ML ;
DOI, K .
MEDICAL PHYSICS, 1984, 11 (03) :287-295
[6]   Performance of a 41X41-cm2 amorphous silicon flat panel x-ray detector for radiographic imaging applications [J].
Granfors, PR ;
Aufrichtig, R .
MEDICAL PHYSICS, 2000, 27 (06) :1324-1331
[7]   Evaluation of an algorithm for the assessment of the MTF using an edge method [J].
Greer, PB ;
van Doorn, T .
MEDICAL PHYSICS, 2000, 27 (09) :2048-2059
[8]   IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751
[9]  
PRATT W.K., 1991, DIGITAL IMAGE PROCES, V2
[10]   CHARACTERIZING DIGITAL IMAGE ACQUISITION DEVICES [J].
REICHENBACH, SE ;
PARK, SK ;
NARAYANSWAMY, R .
OPTICAL ENGINEERING, 1991, 30 (02) :170-177