Middle aortic syndrome treated by stent implantation

被引:40
作者
Brzezinska-Rajszys, G
Qureshi, SA
Ksiazyk, J
Zubrzycka, M
Kosciesza, A
Kubicka, K
Tynan, M
机构
[1] Childrens Mem Hlth Inst, Catheterisat Lab, PL-04736 Warsaw, Poland
[2] Guys Hosp, Dept Paediat Cardiol, London SE1, England
关键词
middle aortic syndrome; stent implantation; paediatric cardiology; interventional cardiology;
D O I
10.1136/hrt.81.2.166
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives-To determine outcome of stent implantation in patients with middle aortic syndrome. Design-Prospective study, case series. Setting-A tertiary paediatric cardiology centre in a children's hospital. Patients-Five patients, aged 4 to 17 years (mean 11.4 years), with upper limb hyper-tension due to middle aortic syndrome. Intervention-Stents were implanted in the mid/lower thoracic/upper abdominal aorta. Main outcome measure-Satisfactory deployment of stents and blood pressure control. Results-In all patients, angiocardiography showed long segment stenosis in the mid or lower thoracic/upper abdominal aorta. The pressure gradient was between 40 and 90 mm Hg (mean 63.2 mm Hg). Seven Palmaz stents were implanted. Immediately after implantation, the gradient decreased to between 0 and 35 mm Hg (mean 13.6 mm Hg). Angiography showed a satisfactory result with widely patent stents in all. In one patient, thrombosis of the stent occurred six days after implantation. This was successfully treated with infusion of alteplase, further balloon dilatation, and implantation of a second stent overlapping the first, both dilated to 10 mm diameter. One patient had elective redilatation of the stent six months after implantation, with further reduction of the gradient from 35 mm Hg to 10 mm Hg. At the latest follow up between three and 20 months (mean 12.2 Methods months) after stent implantation, in four PATIENTS patients blood pressure was better controlled trolled with antihypertensive drugs. One patient was normotensive without drugs. Computed tomography showed no aneurysm formation in the region of the stents. Conclusions-Stent implantation is a preferable alternative to surgery in the treatment of patients with middle aortic syndrome and merits further evaluation.
引用
收藏
页码:166 / 170
页数:5
相关论文
共 24 条
[1]   DETERMINANTS OF HEMODYNAMIC-RESULTS OF BALLOON DILATION OF AORTIC RECOARCTATION [J].
ANJOS, R ;
QURESHI, SA ;
ROSENTHAL, E ;
MURDOCH, I ;
HAYES, A ;
PARSONS, J ;
BAKER, EJ ;
TYNAN, M .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (06) :665-671
[2]   BALLOON-EXPANDABLE STENT REPAIR OF SEVERE COARCTATION OF AORTA [J].
DELEZO, JS ;
PAN, M ;
ROMERO, M ;
MEDINA, A ;
SEGURA, J ;
PAVLOVIC, D ;
MARTINEZ, C ;
TEJERO, I ;
NAVERO, JP ;
TORRES, F ;
LAFUENTE, M ;
HERNANDEZ, E ;
MELIAN, F ;
CONCHA, M .
AMERICAN HEART JOURNAL, 1995, 129 (05) :1002-1008
[3]   BALLOON COARCTATION ANGIOPLASTY IN ADOLESCENTS AND ADULTS - EARLY AND INTERMEDIATE RESULTS [J].
FAWZY, ME ;
DUNN, B ;
GALAL, O ;
WILSON, N ;
SHAIKH, A ;
SRIRAM, R ;
DURAN, CMG .
AMERICAN HEART JOURNAL, 1992, 124 (01) :167-171
[4]   TRANS-LUMINAL BALLOON DILATION OF RESECTED COARCTED SEGMENTS OF THORACIC AORTA - HISTOLOGICAL STUDY AND CLINICAL IMPLICATIONS [J].
HO, SY ;
SOMERVILLE, J ;
YIP, WCL ;
ANDERSON, RH .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1988, 19 (01) :99-105
[5]   ACUTE PSEUDOANEURYSM FORMATION COMPLICATING BALLOON DILATION OF NATIVE COARCTATION - TREATMENT BY DELAYED SURGICAL REPAIR [J].
HUGGON, IC ;
MURDOCH, IA ;
COOKE, AC ;
ANDERSON, DR .
PEDIATRIC CARDIOLOGY, 1994, 15 (06) :313-315
[6]   PSEUDO-ANEURYSM FORMATION FOLLOWING BALLOON ANGIOPLASTY FOR RECURRENT COARCTATION OF THE AORTA [J].
JOYCE, DH ;
MCGRATH, LB .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 20 (02) :133-135
[7]  
Kashani IA, 1996, CATHETER CARDIO DIAG, V38, P406, DOI 10.1002/(SICI)1097-0304(199608)38:4<406::AID-CCD18>3.0.CO
[8]  
2-C
[9]   DISSECTING TRANSVERSE AORTIC-ARCH ANEURYSM AFTER PERCUTANEOUS TRANS-LUMINAL BALLOON DILATION ANGIOPLASTY OF AN AORTIC COARCTATION [J].
KRABILL, KA ;
BASS, JL ;
LUCAS, RV ;
EDWARDS, JE .
PEDIATRIC CARDIOLOGY, 1987, 8 (01) :39-42
[10]   BALLOON ANGIOPLASTY WITH STENT IMPLANTATION IN EXPERIMENTAL COARCTATION OF THE AORTA [J].
MORROW, WR ;
SMITH, VC ;
EHLER, WJ ;
VANDELLEN, AF ;
MULLINS, CE .
CIRCULATION, 1994, 89 (06) :2677-2683