Development and characterisation of a new interface for coupling capillary LC with collision-cell ICP-MS and its application for phosphorylation profiling of tryptic protein digests

被引:55
作者
Pröfrock, D
Leonhard, P
Ruck, W
Prange, A
机构
[1] GKSS Forschungszentrum Geesthacht GmbH, Inst Coastal Res, Marine Bioanalyt Chem, D-21502 Geesthacht, Germany
[2] Univ Luneberg, D-22335 Luneburg, Germany
关键词
capillary-LC; protein; phosphorylation; post-translational modification; ICP-MS; hyphenation; collision cell; reaction cell;
D O I
10.1007/s00216-004-2930-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A comparison of different nebulisers for direct hyphenation of capillary and nano liquid chromatography (Cap-LC, Nano-LC) and quadrupole-based collision cell inductively coupled plasma mass spectrometry (CC-ICP-MS) for phosphorylation profiling of tryptic protein digests is described. Helium was used as cell gas and specially tuned instrumental conditions were used to achieve background minimisation at the mass of phosphorus, because of kinetic energy discrimination of the interfering polyatomic ions. The proposed set-up is based on a modified capillary electrophoresis interface and a home-made 4-mL spray chamber. It enables the use of gradient conditions with a highly concentrated organic mobile phase as often used in protein phosphorylation analysis, without the need to apply membrane desolvation for removal of the organic phase or further background minimisation. No significant signal suppression or other negative effects caused by the organic mobile phase occur, because of the low flow rates used in Cap-LC and the robust plasma conditions of the CC-ICP-MS instrument. A tryptic digest of beta-casein was investigated as model compound to demonstrate the applicability of the proposed set-up for phosphorylation profiling in protein analysis using quadrupole based collision-cell ICP-MS as phosphorus-specific detector. Detection limits for phosphorylated peptides down to the sub picomole level were obtained. As a complementary technique, electrospray ionisation tandem mass spectrometry (ESI-MS-MS) with data base searching was used for further characterisation of the phosphorylated peptides detected.
引用
收藏
页码:194 / 204
页数:11
相关论文
共 35 条
[1]   Identification of phosphopeptides by chemical modification with an isotopic tag and ion trap mass spectrometry [J].
Adamczyk, M ;
Gebler, JC ;
Wu, J .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (10) :999-1001
[3]   Sequencing of gel-isolated proteins using microblotter capillary liquid chromatography-electrospray mass spectrometry [J].
Bauer, MD ;
Sun, YP ;
Wang, F .
JOURNAL OF PROTEIN CHEMISTRY, 1999, 18 (03) :337-347
[4]   Phosphopeptide detection and sequencing by matrix-assisted laser desorption/ionization quadrupole time-of-flight tandem mass spectrometry [J].
Bennett, KL ;
Stensballe, A ;
Podtelejnikov, AV ;
Moniatte, M ;
Jensen, ON .
JOURNAL OF MASS SPECTROMETRY, 2002, 37 (02) :179-190
[5]   De novo peptide sequencing and quantitative profiling of complex protein mixtures using mass-coded abundance tagging [J].
Cagney, G ;
Emili, A .
NATURE BIOTECHNOLOGY, 2002, 20 (02) :163-170
[6]   Phosphoproteome analysis by mass spectrometry and its application to Saccharomyces cerevisiae [J].
Ficarro, SB ;
McCleland, ML ;
Stukenberg, PT ;
Burke, DJ ;
Ross, MM ;
Shabanowitz, J ;
Hunt, DF ;
White, FM .
NATURE BIOTECHNOLOGY, 2002, 20 (03) :301-305
[7]   Selective, sensitive, and rapid phosphopeptide identification in enzymatic digests using ESI-FTICR-MS with infrared multiphoton dissociation [J].
Flora, JW ;
Muddiman, DC .
ANALYTICAL CHEMISTRY, 2001, 73 (14) :3305-3311
[8]   PACKED-CAPILLARY LIQUID-CHROMATOGRAPHY ION-SPRAY TANDEM MASS-SPECTROMETRY DETERMINATION OF BIOMOLECULES [J].
HUANG, EC ;
HENION, JD .
ANALYTICAL CHEMISTRY, 1991, 63 (07) :732-739
[9]  
Lee CH, 2001, RAPID COMMUN MASS SP, V15, P191, DOI 10.1002/1097-0231(20010215)15:3<191::AID-RCM209>3.0.CO
[10]  
2-N