Consideration of the uncertainty in the CRM value with the analysis of solid materials using atomic spectrometry - Part 1. Evaluation and benefit of method uncertainty using certified reference materials

被引:5
作者
Kurfurst, U [1 ]
机构
[1] Univ Appl Sci, Fachhsch Fulda, Dept Housing & Nutr, D-36039 Fulda, Germany
关键词
uncertainty; atomic spectrometry; certified reference material; method validation; trueness;
D O I
10.1007/s007690050274
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The principles of the ISO "Guide to the expression of uncertainty in measurement" are applied to the analysis of solid materials by graphite furnace atomic absorption spectrometry. Assessment of trueness is achieved by the analysis of a certified reference material (CRM) under fixed instrumental conditions. For the evaluation of the method's uncertainty las part of the validation) a model equation is derived, considering all significant contributions, including the uncertainty in the CRM value and the uncertainty in the CRM analysis. A concrete example is presented and discussed (soya flour as an unknown sample and CRM BCR rye grass). For routine analysis a pooled estimate from the validation experiment can be used, leading to an acceptable small uncertainty range even in the case of "duplicate determination". The test for trueness (acceptance) of the result of the CRM analysis is applied and the correction for detected bias is discussed.
引用
收藏
页码:406 / 411
页数:6
相关论文
共 11 条
[1]  
*EURACHEM, 1998, NEWSL, V13, P4
[2]  
*EURACHEM, 1995, QUANT UNC AN MEAS
[3]  
GRIEPINK B, 1988, EUR11839EN BCR INF C
[4]   Derived measurement standards of reduced uncertainty - A contradiction? [J].
Hasselbarth, W ;
Bremser, W .
ACCREDITATION AND QUALITY ASSURANCE, 1998, 3 (08) :337-339
[5]  
International Organization for Standardization, 1993, GUID EXPR UNC MEAS
[6]  
KESSEL R, 1997, GUM WORKBENCH VERSIO
[7]   Uncertainty in analytical results from solid materials with electrothermal atomic absorption spectrometry: A comparison methods [J].
Kurfurst, U ;
Rehnert, A ;
Muntau, H .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (02) :229-244
[8]  
KURFURST U, 1995, GIT FACHZ LAB, V7, P662
[9]  
KURFURST U, 1998, SOLID SAMPLE ANAL DI
[10]   A view of uncertainty at the bench analytical level [J].
Rios, A ;
Valcarcel, M .
ACCREDITATION AND QUALITY ASSURANCE, 1998, 3 (01) :14-19