Evaluation of an IgM immunoblot kit for dengue diagnosis

被引:22
作者
Kuno, G
Cropp, CB
Wong-Lee, J
Gubler, DJ
机构
[1] Ctr Dis Control & Prevent, Div Vector Borne Infect Dis, Natl Ctr Infect Dis, Ft Collins, CO 80525 USA
[2] Genelabs Technol Inc, Redwood City, CA USA
关键词
D O I
10.4269/ajtmh.1998.59.757
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
A commercial IgM immunoblot kit was evaluated for dengue diagnosis with a panel of serum specimens collected from patients in a dengue endemic area. The kit is not recommended for use in its present form because of its undesirable rate of false-positive results. However, by substituting internal controls with the reference positive and negative controls that are more representative of those seen in endemic areas and by modifying the positive and negative scoring criteria, sensitivity and specificity of 80.3% and 94.5%, respectively, were obtained. These results are comparable with those obtained with the IgM ELISA on specimens, most of which were obtained from outpatient health care facilities. With further technical modifications, inclusion of a visual guide to ensure scoring standardization, and a more complete elaboration of the limitations of the test, wide application of the kit in diagnostic laboratories should be possible.