Comparative evaluation of the QUANTIPLEX HIV-1 RNA 2.0 and 3.0 (bDNA) assays and the AMPLICOR HIV-1 MONITOR v1.5 test for the quantitation of human immunodeficiency virus type 1 RNA in plasma

被引:16
作者
Anastassopoulou, CG
Toulomi, G
Katsoulidou, A
Hatzitheodorou, H
Pappa, M
Paraskevis, D
Lazanas, M
Gargalianos, P
Hatzakis, A
机构
[1] Univ Athens, Sch Med, Natl Retrovirus Reference Ctr, Dept Hyg & Epidemiol, GR-11527 Athens, Greece
[2] Tzanio Gen Hosp, Piraeus, Greece
[3] Gen Hosp Athens G Gennimatas, Athens, Greece
关键词
HIV-1; RNA; viral load; branched DNA; Amplicor;
D O I
10.1016/S0166-0934(00)00245-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
HIV-1 RNA measurements from 84 plasma specimens obtained with the QUANTIPLEX HIV-1 RNA 2.0 and 3.0 (bDNA) assays (Chiron Diagnostics, Emeryville, CA) and with the AMPLICOR HIV-1 MONITOR (TM) Test, version 1.5 with ultra-sensitive specimen preparation (Roche Diagnostic Systems, Inc., Branchburg, NJ) were compared. The absolute RNA values of tested specimens differed significantly between bDNA 2.0 and bDNA 3.0 or Monitor v1.5 measurements (Wilcoxon signed-rank test P < 0.001). Results generated with bDNA 3.0 or with Monitor v1.5 were approximately twofold greater than those generated with bDNA 2.0, with smaller differences at higher HIV-1 RNA levels and greater differences at RNA levels below 1000 copies per mi. Although highly correlated (r = 0.92 and 0.86, respectively), viral load data generated with bDNA 2.0 and either bDNA 3.0 or Monitor v1.5 were in poor agreement. Concordant results (difference in log(10) copies per ml < 0.5) were found at frequencies of 80% for bDNA 2.0 and bDNA 3.0 and only at 58.5% for bDNA 2.0 and Monitor v1.5. In contrast, bDNA 3.0 and Monitor v1.5 measurements were highly correlated (r = 0.96) and in good agreement (92.7%). (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:67 / 74
页数:8
相关论文
共 19 条
  • [1] Aschbacher R, 1999, MICROBIOLOGICA, V22, P1
  • [2] *CDCP, 1998, MMWR-MORBID MORTAL W, V47, P43
  • [3] CLARK JR, 1999, 7 EUR C CLIN ASP TRE
  • [4] A branched DNA signal amplification assay for quantification of nucleic acid targets below 100 molecules/ml
    Collins, ML
    Irvine, B
    Tyner, D
    Fine, E
    Zayati, C
    Chang, CA
    Horn, T
    Ahle, D
    Detmer, J
    Shen, LP
    Kolberg, J
    Bushnell, S
    Urdea, MS
    Ho, DD
    [J]. NUCLEIC ACIDS RESEARCH, 1997, 25 (15) : 2979 - 2984
  • [5] Evaluation of the ultrasensitive Roche Amplicor HIV-1 monitor assay for quantitation of human immunodeficiency virus type I RNA
    Erali, M
    Hillyard, DR
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (03) : 792 - 795
  • [6] Fischl MA, 1999, AIDS, V13, pS49
  • [7] Comparison of the Quantiplex version 3.0 assay and a sensitized Amplicor Monitor assay for measurement of human immunodeficiency virus type 1 RNA levels in plasma samples
    Highbarger, HC
    Alvord, WG
    Jiang, MK
    Shah, AS
    Metcalf, JA
    Lane, HC
    Dewar, RL
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (11) : 3612 - 3614
  • [8] KATSENSTEIN DA, 1996, NEW ENGL J MED, V335, P1091
  • [9] Prognosis in HIV-1 infection predicted by the quantity of virus in plasma
    Mellors, JW
    Rinaldo, CR
    Gupta, P
    White, RM
    Todd, JA
    Kingsley, LA
    [J]. SCIENCE, 1996, 272 (5265) : 1167 - 1170
  • [10] Plasma viral load and CD4(+) lymphocytes as prognostic markers of HIV-1 infection
    Mellors, JW
    Munoz, A
    Giorgi, JV
    Margolick, JB
    Tassoni, CJ
    Gupta, P
    Kingsley, LA
    Todd, JA
    Saah, AJ
    Detels, R
    Phair, JP
    Rinaldo, CR