共 28 条
Arsenic determination in saline waters utilizing a tubular membrane as a gas-liquid separator for hydride generation inductively coupled plasma mass spectrometry
被引:26
作者:
Creed, JT
Magnuson, ML
Brockhoff, CA
Chamberlain, I
Sivaganesan, M
机构:
[1] US EPA,REG 10,MANCHESTER LAB,PORT ORCHARD,WA
[2] DYNCORP,CINCINNATI,OH 45219
关键词:
hydride generation;
membrane-based gas-liquid separator;
inductively coupled plasma mass spectrometry;
D O I:
10.1039/ja9961100505
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A tubular silicone rubber membrane is evaluated as a gas-liquid separator for the determination of arsenic in saline waters aia HG-ICP-MS. The system was optimized in terms of NaBH4 and HCl concentrations. The intermediate gas and carrier gas were optimized in terms of sensitivity and stability. The pre-reduction of As-V in standard solutions and reference materials was investigated using KI and KI plus ascorbic acid sample treatments. The KI plus ascorbic acid treatment produced a similar response for Asm and As-V independent of the matrix. The KI plus ascorbic acid treated samples could be held for 21 days without significant response changes. The detection limit was 92 pg based on 7 replicate injections of a 400 pg standard. The overall system was evaluated by analysing two reference materials, NASS-4 and SLEW-2. Precision and recovery data were collected on three estuarine samples.
引用
收藏
页码:505 / 509
页数:5