Trace analysis of sulfonylurea herbicides in water by on-line continuous flow liquid membrane extraction-C18 precolumn liquid chromatography with ultraviolet absorbance detection

被引:47
作者
Liu, JF
Chao, JB
Jiang, GB
Cai, YQ
Liu, JM
机构
[1] Chinese Acad Sci, Ecoenvironm Sci Res Ctr, Key Lab Environm Chem & Ecotoxicol, Beijing 100085, Peoples R China
[2] Beijing Univ Sci & Technol, Dept Chem, Beijing 100083, Peoples R China
基金
中国国家自然科学基金;
关键词
trace analysis; water analysis; continuous-flow liquid membrane extraction; sulfonylureas; pesticides;
D O I
10.1016/S0021-9673(03)00515-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An on-line system that consists of continuous-flow liquid membrane extraction (CFLME), C-18 precolumn, and liquid chromatography with UV detection was applied to trace analysis of sulfonylurea herbicides in water. During preconcentration by CFLME, five target compounds, including metsulfuron methyl, bensulfuron methyl, tribenuron methyl, sulfometuron methyl, and ethametsulfuron, were enriched in 960 mul of 0.5 mol l(-1) Na2CO3-NaHCO3 (pH 10.8) buffer used as acceptor. This acceptor was on-line neutralized and transported to the C-18 precolumn where the analytes were absorbed and focused. Then the focused analytes were injected onto a C-18 analytical column for separation and detection at 240 nm. The proposed method was applied to determine sulfonylurea herbicides in water, river, and reservoir water with detection limits of 10-50 ng l(-1) when enriching a 120-ml sample. Throughput is typically one sample per hour. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:21 / 28
页数:8
相关论文
共 12 条