On the effects of different strategies in modelling balloon-expandable stenting by means of finite element method

被引:174
作者
Gervaso, Francesca [1 ]
Capelli, Claudio [1 ]
Petrini, Lorenza [1 ]
Lattanzio, Simone [1 ]
Di Virgilio, Luca [1 ]
Migliavacca, Francesco [1 ]
机构
[1] Politecn Milan, Dept Struct Engn, Lab Biol Struct Mech, I-20133 Milan, Italy
关键词
balloon-expandable stent; finite element method; mathematical model;
D O I
10.1016/j.jbiomech.2008.01.027
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
In recent years, computational structural analyses have emerged as important tools to investigate the mechanical response of stent placement into arterial walls. Although most coronary stents are expanded by inflating a polymeric balloon, realistic computational balloon models have been introduced only recently. In the present study, the finite element method is applied to simulate three different approaches to evaluate stent-free expansion and stent expansion inside an artery. Three different stent expansion modelling techniques were analysed by: (i) imposing a uniform pressure on the stent internal surface, (ii) a rigid cylindrical surface expanded with displacement control and (iii) modelling a polymeric deformable balloon. The computational results showed differences in the free and confined-stent expansions due to different expansion techniques. The modelling technique of the balloon seems essential to estimate the level of injury caused on arterial walls during stent expansion. (c) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1206 / 1212
页数:7
相关论文
共 36 条
[1]
Auricchio F., 2001, COMPUT METHOD BIOMEC, V4, P249, DOI DOI 10.1080/10255840108908007
[2]
Intramural stress increases exponentially with stent diameter: A stress threshold for neointimal hyperplasia [J].
Ballyk, Peter D. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (07) :1139-1145
[3]
Effects of stent design parameters on normal artery wall mechanics [J].
Bedoya, Julian ;
Meyer, Clark A. ;
Timmins, Lucas H. ;
Moreno, Michael R. ;
Moore, James E., Jr. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2006, 128 (05) :757-765
[4]
De Beule M, 2006, Technol Health Care, V14, P233
[5]
DEBEULE M, 2007, IN PRESS J BIOMECHAN, DOI DOI 10.1016/JJBIOMECH.2007.08.014
[6]
Mechanical behaviour modelling of balloon-expandable stents [J].
Dumoulin, C ;
Cochelin, B .
JOURNAL OF BIOMECHANICS, 2000, 33 (11) :1461-1470
[7]
Pathobiologic responses to stenting [J].
Edelman, ER ;
Rogers, C .
AMERICAN JOURNAL OF CARDIOLOGY, 1998, 81 (7A) :4E-6E
[8]
Finite element modeling of balloon angioplasty by considering overstretch of remnant non-diseased tissues in lesions [J].
Gasser, T. Christian ;
Holzapfel, Gerhard A. .
COMPUTATIONAL MECHANICS, 2007, 40 (01) :47-60
[9]
The finite element analysis of stresses in atherosclerotic arteries during balloon angioplasty [J].
Gourisankaran, V ;
Sharma, MG .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 2000, 28 (1-2) :47-51
[10]
Comparison of element technologies for modeling stent expansion [J].
Hall, Garrett J. ;
Kasper, Eric P. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2006, 128 (05) :751-756