Implementing new recommendations for calibrating personal dosemeters

被引:4
作者
Ginjaume, M [1 ]
Ortega, X [1 ]
Duch, MA [1 ]
机构
[1] Tech Univ Catalonia, Inst Energy Technol, Barcelona 08028, Spain
关键词
D O I
10.1093/oxfordjournals.rpd.a006615
中图分类号
X [环境科学、安全科学];
学科分类号
08 [工学]; 0830 [环境科学与工程];
摘要
This paper analyses the differences between the calibration procedures for personal dosemeters recommended by ICRU 47 and ISO 4037-3. The tissue equivalence of the PMMA and the ISO water slab phantoms are analysed by means Of the Penelope Monte Carlo code for monoenergetic and filtered X ray photon beams and compared with the results of two other independent codes. The influence of the calibration method is also verified experimentally, both on a thermoluminescence and an electronic personal dosemeter. Good consistency between both calibration procedures is shown provided that a correction factor for backscatter differences between the PMMA and the ICRU phantom is introduced. The Monte Carlo simulation is used to determine this correction to a greater accuracy.