Calibration curves for real-time PCR

被引:17
作者
Lai, KKY
Cook, L
Krantz, EM
Corey, L
Jerome, KR
机构
[1] Fred Hutchinson Canc Res Ctr, Program Infect Dis, Seattle, WA 98109 USA
[2] Univ Washington, Med Ctr, Dept Lab Med, Seattle, WA 98195 USA
[3] Univ Washington, Med Ctr, Dept Med, Seattle, WA 98195 USA
关键词
D O I
10.1373/clinchem.2004.039909
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Despite the increasing use of real-time PCR in the diagnosis and management of viral infections, there are no published studies adequately addressing the optimum number of calibrators, the number of replicates of each calibrator, and the frequency with which calibration needs to be repeated. This study was designed to address these issues. Methods: Cycle threshold data (ABI 7700) was collected from > 50 consecutive real-time PCR runs for hepatitis B and Epstein-Barr viruses. Our routine calibration curve made from serial 10-fold dilutions run in duplicate was compared with alternative options, including duplicate 100-fold dilutions, inclusion of a low-copy calibrator, and omission of the duplicate determination. Control data were used to examine the use of an average calibration curve made from multiple runs. Results: Use of duplicate serial 10-fold dilutions led to the least imprecision, duplicate 100-fold dilutions had slightly higher imprecision, and calibration curves obtained with singlet measurements showed the greatest imprecision. For patient data, the duplicate 100-fold dilution calibration curve produced results that best matched those from the routine calibration curve. Use of singlet dilutions or inclusion of a low-copy calibrator produced poorer agreement. Variability in controls was lower with a daily calibration curve than with an average calibration curve. Conclusions: Duplicate 100-fold dilution calibration curves produced equivalent results and the same imprecision as curves with more calibrators, and thus are a valid alternative. Laboratories should carefully evaluate the variability resulting from the use of average calibration curves before adopting this approach. (c) 2005 American Association for Clinical Chemistry
引用
收藏
页码:1132 / 1136
页数:5
相关论文
共 9 条
[1]   Quantitation of hepatitis B virus genomic DNA by real-time detection PCR [J].
Abe, A ;
Inoue, K ;
Tanaka, T ;
Kato, J ;
Kajiyama, N ;
Kawaguchi, R ;
Tanaka, S ;
Yoshiba, M ;
Kohara, M .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (09) :2899-2903
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   ROBUST LOCALLY WEIGHTED REGRESSION AND SMOOTHING SCATTERPLOTS [J].
CLEVELAND, WS .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1979, 74 (368) :829-836
[4]   Detection of Epstein-Barr virus DNA in sera from transplant recipients with lymphoproliferative disorders [J].
Limaye, AP ;
Huang, ML ;
Atienza, EE ;
Ferrenberg, JM ;
Corey, L .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (04) :1113-1116
[5]   Real-time quantitative PCR for human herpesvirus 6 DNA [J].
Locatelli, G ;
Santoro, F ;
Veglia, F ;
Gobbi, A ;
Lusso, P ;
Malnati, MS .
JOURNAL OF CLINICAL MICROBIOLOGY, 2000, 38 (11) :4042-4048
[6]   High-throughput quantitative analysis of hepatitis B virus DNA in serum using the TaqMan fluorogenic detection system [J].
Loeb, KR ;
Jerome, KR ;
Goddard, J ;
Huang, ML ;
Cent, A ;
Corey, L .
HEPATOLOGY, 2000, 32 (03) :626-629
[7]   Real-time PCR in virology [J].
Mackay, IM ;
Arden, KE ;
Nitsche, A .
NUCLEIC ACIDS RESEARCH, 2002, 30 (06) :1292-1305
[8]  
NETER J, 1996, APPL LINEAR STAT MOD, P67
[9]   Development of a real-time quantitative RT-PCR for the detection of HIV-2 RNA in plasma [J].
Schutten, M ;
van den Hoogen, B ;
van der Ende, ME ;
Gruters, RA ;
Osterhaus, ADME ;
Niesters, HGM .
JOURNAL OF VIROLOGICAL METHODS, 2000, 88 (01) :81-87