Ascending aorta outflow graft location and pulsatile ventricular assist provide optimal hemodynamic support in an adult mock circulation

被引:25
作者
Litwak, KN [1 ]
Koenig, SC [1 ]
Cheng, RC [1 ]
Giridharan, GA [1 ]
Gillars, KJ [1 ]
Pantalos, GM [1 ]
机构
[1] Univ Louisville, Jewish Hosp Heart & Lung Inst, Dept Surg, Louisville, KY 40292 USA
关键词
ventricular assist; mock circulation; hemodynamics;
D O I
10.1111/j.1525-1594.2005.29100.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Although continuous flow (CF) and pulsatile flow (PF) ventricular assist devices (VADs) are being clinically used, their effects on aortic blood flow, as a measure of overall blood distribution, remain unclear. In acute VAD support animal experiments, our group has described a zone of turbulent mixing in the aortic arch. The objective of this study was to confirm this finding in the controlled setting of an adult mock circulation, simulating ventricular pathophysiologic states (normal and failing ventricle). CF and PF flow VADs were connected to ventricular apical inflow and ascending aorta (AA) or descending aorta (DA) outflow cannulae. Cardiovascular pressure and flow waveforms were recorded at varying levels of VAD bypass resulting in four test conditions: (i) CF-AA; (ii) CF-DA; (iii) PF-AA; and (iv) PF-DA. Confirming the animal data, no differences in mean aortic flow between CF and PF VADs were found, and significantly lower mean aortic arch flow with DA cannulation was noted. Mean aortic root flow decreased with increasing VAD bypass flow. As in the animal studies, despite similar mean flow rates, significant differences in waveform morphology were observed for AA and DA outflow graft locations and varying levels of VAD bypass. At 100% VAD support in the failing heart, PF restored waveform pulsatility to normal baseline while CF resulted in little pulsatility. These results confirm our earlier findings in the animal model, suggesting that outflow graft location may have a significant effect on aortic blood flow distribution. The long-term implications of these findings are being examined in ongoing studies.
引用
收藏
页码:629 / 635
页数:7
相关论文
共 24 条
[1]   First successful Italian clinical experience with DeBakey VAD™ [J].
Agati, S ;
Bruschi, G ;
Russo, C ;
Colombo, T ;
Lanfranconi, M ;
Vitali, E .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2001, 20 (08) :914-917
[2]  
DiGiorgi P. L., 2002, Journal of Heart and Lung Transplantation, V21, P151, DOI 10.1016/S1053-2498(01)00703-3
[3]  
DREW GA, 2000, LABVIEW AUTOMATIVE T, P92
[4]   Research and development of an implantable, axial-flow left ventricular assist device: The Jarvik 2000 heart [J].
Frazier, OH ;
Myers, TJ ;
Jarvik, RK ;
Westaby, S ;
Pigott, DW ;
Gregoric, ID ;
Khan, T ;
Tamez, DW ;
Conger, JL ;
Macris, MP .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S125-S132
[5]   Left ventricular and myocardial perfusion responses to volume unloading and afterload reduction in a computer simulation [J].
Giridharan, GA ;
Ewert, DL ;
Pantalos, GM ;
Gillars, KJ ;
Litwak, KN ;
Gray, LA ;
Koenig, SC .
ASAIO JOURNAL, 2004, 50 (05) :512-518
[6]  
Goldstein Sidney, 1998, Cardiology Clinics, V16, P623, DOI 10.1016/S0733-8651(05)70039-4
[7]  
Griffith BP, 2001, ANN THORAC SURG, V71, pS114
[8]   Pulsatile and nonpulsatile perfusion: The continuing controversy [J].
Hornick, P ;
Taylor, K .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1997, 11 (03) :310-315
[9]   Smooth muscle cell hypertrophy of renal cortex arteries with chronic continuous flow left ventricular assist [J].
Kihara, S ;
Litwak, KN ;
Nichols, L ;
Litwak, P ;
Kameneva, MV ;
Wu, ZJ ;
Kormos, RL ;
Griffith, BP .
ANNALS OF THORACIC SURGERY, 2003, 75 (01) :178-183
[10]   Mechanical circulatory support devices in the treatment of heart failure [J].
Knight, JL .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF MEDICINE, 1999, 29 (03) :410-417