Design considerations to overcome cross talk in a photon counting silicon strip detector for computed tomography

被引:25
作者
Bornefalk, Hans [1 ]
Xu, Cheng [1 ]
Svensson, Christer [2 ]
Danielsson, Mats [1 ]
机构
[1] Royal Inst Technol, Dept Phys, SE-10691 Stockholm, Sweden
[2] Linkoping Univ, Div Elect Devices, SE-58183 Linkoping, Sweden
关键词
Monte Carlo simulation; Silicon microstrip detector; Photon counting CT; Cross talk; Pileup; RAY; DEPENDENCE;
D O I
10.1016/j.nima.2010.04.071
中图分类号
TH7 [仪器、仪表];
学科分类号
080401 [精密仪器及机械];
摘要
This article presents a Monte Carlo simulation of the detector energy response in the presence of pileup in a segmented silicon microstrip detector designed for high flux spectral computed tomography with sub-millimeter pixel size. Currents induced on the collection electrode of a pixel segment are explicitly modeled and signals emanating from events in neighboring pixels are superimposed together with electronic noise before the entire pulse train is processed by a model of the readout electronics to obtain the detector energy response function. The article shows how the lower threshold and the time constant of the electronic filters need to be set in order to minimize the detrimental influence of cross talk from neighboring pixel segments, an issue that is aggravated by the sub-millimeter pixel size and the proposed segmented detector design. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:371 / 378
页数:8
相关论文
共 23 条
[1]
[Anonymous], P SPIE
[2]
Berger M J., 1999, XCOM: photon cross sections database
[3]
Photon-counting spectral computed tomography using silicon strip detectors: a feasibility study [J].
Bornefalk, Hans ;
Danielsson, Mats .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (07) :1999-2022
[4]
A new Monte Carlo code for full simulation of silicon strip detectors [J].
Brigida, M ;
Favuzzi, C ;
Fusco, P ;
Gargano, F ;
Giglietto, N ;
Giordano, F ;
Loparco, F ;
Marangelli, B ;
Mazziotta, MN ;
Mirizzi, N ;
Rainò, S ;
Spinelli, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 533 (03) :322-343
[5]
High-resolution X-ray imaging using the signal time dependence on a double-sided silicon detector [J].
Cederström, B ;
Danielsson, M ;
Lundqvist, M ;
Nygren, D .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 423 (01) :135-145
[6]
CRANLEY K, 1997, 78 I PHYS ENG MED DI
[7]
ELECTRON PENETRATION AND ENERGY-TRANSFER IN SOLID TARGETS [J].
FITTING, HJ ;
GLAEFEKE, H ;
WILD, W .
PHYSICA STATUS SOLIDI A-APPLIED RESEARCH, 1977, 43 (01) :185-190
[8]
CdZnTe detector array for a Scanning-Beam Digital-X-ray system [J].
Heanue, JA ;
Pearson, DA ;
Melen, RE .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :718-725
[9]
TEMPERATURE-DEPENDENCE OF HOLE SATURATION VELOCITY IN SILICON [J].
IKOMA, T ;
HARA, K .
APPLIED PHYSICS, 1974, 3 (05) :431-432
[10]
About the Scattering of Radiation by free Electrons after the new relativistic Quantum dynamics of Dirac. [J].
Klein, O. ;
Nishina, Y. .
ZEITSCHRIFT FUR PHYSIK, 1929, 52 (11-12) :853-868