Evaluation of key sources of variability in the measurement of pharmaceutical drug products by near infrared reflectance spectroscopy

被引:19
作者
Borer, MW [1 ]
Zhou, XJ [1 ]
Hays, DM [1 ]
Hofer, JD [1 ]
White, KC [1 ]
机构
[1] Eli Lilly & Co, Lilly Res Labs, Indianapolis, IN 46285 USA
关键词
near infrared; NIR; NIRA; NIRS; reflectance spectroscopy; clinical trial materials; experimental design; product discrimination; identification test;
D O I
10.1016/S0731-7085(97)00229-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Potential sources of variability in the measurement of solid oral drug products by near infrared reflectance spectroscopy were evaluated with statistical experimental design. Spectra were collected for two different tablet types according to the data collection and treatment parameters defined by the experimental design. Each tablet had three different dose-levels. Libraries were constructed using second-derivative spectra. Key figures-of-merit generated during internal and external library validation were used to calculate which parameters most strongly influence the library performance for dose-level discrimination. These responses and their corresponding experimental conditions were evaluated with the screening model in the JMP(R) program. Segment value used for the second-derivative calculation was an influential factor and had a complex effect. Orientation on the sampling platform also had an influential effect for embossed tablets. Collection of spectra over fewer days decreased variability within the library. More frequent reference spectrum collection improved the performance of libraries to a small degree. A larger sample population increased the range of spectral variability within a dose-level but apparently not the overall performance of the library. The number of scans averaged per spectrum was not an influential factor in this study. These results are summarized and used to recommend an approach to dose-level discrimination. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:641 / 650
页数:10
相关论文
共 13 条
[1]   IDENTIFICATION OF TABLET FORMULATIONS INSIDE BLISTER PACKAGES BY NEAR-INFRARED SPECTROSCOPY [J].
ALDRIDGE, PK ;
MUSHINSKY, RF ;
ANDINO, MM ;
EVANS, CL .
APPLIED SPECTROSCOPY, 1994, 48 (10) :1272-1276
[2]   DETERMINATION OF ASCORBIC-ACID IN PHARMACEUTICAL PREPARATIONS BY NEAR-INFRARED REFLECTANCE SPECTROSCOPY [J].
BLANCO, M ;
COELLO, J ;
ITURRIAGA, H ;
MASPOCH, S ;
DELAPEZUELA, C .
TALANTA, 1993, 40 (11) :1671-1676
[3]  
BUICE RG, 1995, PHARMACEUT RES, V12, P161
[4]   USES OF NEAR-INFRARED SPECTROSCOPY IN PHARMACEUTICAL ANALYSIS [J].
CIURCZAK, EW .
APPLIED SPECTROSCOPY REVIEWS, 1987, 23 (1-2) :147-163
[5]  
CORTI P, 1992, PROCESS CONTR QUAL, V2, P131
[6]   NEAR-INFRARED METHODS FOR THE IDENTIFICATION OF TABLETS IN CLINICAL-TRIAL SUPPLIES [J].
DEMPSTER, MA ;
JONES, JA ;
LAST, IR ;
MACDONALD, BF ;
PREBBLE, KA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1993, 11 (11-12) :1087-1092
[7]   QUALITY-CONTROL IN MANUFACTURING PROCESS BY NEAR-INFRARED SPECTROSCOPY [J].
GONZALEZ, F ;
POUS, R .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1995, 13 (4-5) :419-423
[8]   NEAR-INFRARED SPECTROSCOPY - APPLICATIONS IN THE ANALYSIS OF TABLETS AND SOLID PHARMACEUTICAL DOSAGE FORMS [J].
KIRSCH, JD ;
DRENNEN, JK .
APPLIED SPECTROSCOPY REVIEWS, 1995, 30 (03) :139-174
[9]   SUITABILITY OF NEAR-INFRARED METHODS FOR THE DETERMINATION OF MOISTURE IN A FREEZE-DRIED INJECTION PRODUCT CONTAINING DIFFERENT AMOUNTS OF THE ACTIVE INGREDIENT [J].
LAST, IR ;
PREBBLE, KA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1993, 11 (11-12) :1071-1076
[10]   ANALYSIS OF INTACT TABLETS BY NEAR-INFRARED REFLECTANCE SPECTROMETRY [J].
LODDER, RA ;