A wireless implantable multichannel micro stimulating system-on-a-chip with modular architecture

被引:106
作者
Ghovanloo, Maysam [1 ]
Najafi, Khalil
机构
[1] Georgia Inst Technol, Dept Elect & Comp Engn, GT Bion Lab, Atlanta, GA 30332 USA
[2] Univ Michigan, Ctr Wireless Integrated Microsyst, Ann Arbor, MI 48109 USA
关键词
charge balancing; frequency shift keying; implant; microstimulation; modular architecture; neural prostheses; stimulation strategy; system-on-a-chip (SoP); wireless link;
D O I
10.1109/TNSRE.2007.903970
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A 64-site wireless current microstimulator chip (Interestim-2B) and a prototype implant based on the same chip have been developed for neural prosthetic applications. Modular standalone architecture. allows up to 32 chips to be individually addressed and operated in parallel to drive up to 2048 stimulating sites. The only off-chip components are a receiver inductive-capacitive (LC) tank, a capacitive low-pass filter for ripple rejection, and arrays of microelectrodes for interfacing with the neural tissue. The implant receives inductive power up to 50 mW and data at 2.5 Mb/s from a frequency shift keyed (FSK) 5110 MHZ carrier to generate up to 65 800 stimulus pulses/s. Each Interestim-2B chip contains 16 current drivers with 270 mu A full-scale current, 5-bit (32-steps) digital-to-analog converter (DAC) resolution, 100 M Omega output impedance, and a voltage compliance that extends within 150 and 250 mV of the 5 V supply and ground rails, respectively. It can generate any arbitrary current waveform and supports a variety of monopolar and bipolar stimulation protocols. A common analog line provides access to each site potential, and exhausts residual stimulus charges for charge balancing. The chip has site potential measurement and in situ site impedance measurement capabilities, which help its users indicate defective sites or characteristic shifts in chronic stimulations. Interestim-2B chip is fabricated in the AMI 1.5 mu m standard complementary metal-oxide-semiconductor (CMOS) process and measures 4.6 x 4.6 x 0.5 mm. The prototype implant size including test connectors is 19 x 14 x 6 mm, which can be shrunk down to < 0.5 CC. This paper also summarizes some of the in vitro and in vivo experiments performed using the Interestim-2B prototype implant.
引用
收藏
页码:449 / 457
页数:9
相关论文
共 42 条
[1]  
[Anonymous], WILEY ENCY BIOMEDICA
[2]   CHARGE INJECTION LIMITS OF ACTIVATED IRIDIUM OXIDE ELECTRODES WITH 0.2MS PULSES IN BICARBONATE BUFFERED SALINE [J].
BEEBE, X ;
ROSE, TL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (06) :494-495
[3]   Therapeutic electrical stimulation of the central nervous system [J].
Benabid, AL ;
Wallace, B ;
Mitrofanis, J ;
Xia, C ;
Piallat, B ;
Fraix, V ;
Batir, A ;
Krack, P ;
Pollak, P ;
Berger, F .
COMPTES RENDUS BIOLOGIES, 2005, 328 (02) :177-186
[4]   Deep brain stimulation [J].
Breit, S ;
Schulz, JB ;
Benabid, AL .
CELL AND TISSUE RESEARCH, 2004, 318 (01) :275-288
[5]   Potential-biased, asymmetric waveforms for charge-injection with activated iridium oxide (AIROF) neural stimulation electrodes [J].
Cogan, SF ;
Troyk, PR ;
Ehrlich, J ;
Plante, TD ;
Detlefsen, DE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (02) :327-332
[6]   An arbitrary waveform stimulus circuit for visual prostheses using a low-area multibias DAC [J].
DeMarco, SC ;
Liu, WT ;
Singh, PR ;
Lazzi, G ;
Humayun, MS ;
Weiland, JD .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2003, 38 (10) :1679-1690
[7]  
DOMMEL N, 2006, P 28 IEEE EMBS C NEW, P4651
[8]   THE MOTOR CORTEX OF THE RAT - CYTOARCHITECTURE AND MICROSTIMULATION MAPPING [J].
DONOGHUE, JP ;
WISE, SP .
JOURNAL OF COMPARATIVE NEUROLOGY, 1982, 212 (01) :76-88
[9]   Fully integrated wideband high-current rectifiers for inductively powered devices [J].
Ghovanloo, M ;
Najafi, K .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2004, 39 (11) :1976-1984
[10]   A modular 32-site wireless neural stimulation microsystem [J].
Ghovanloo, M ;
Najafi, K .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2004, 39 (12) :2457-2466