Comparison of blood flow rates and hydraulic resistance between the Mahurkar catheter, the Tesio twin catheter, and the Ash Split Cath

被引:32
作者
Mankus, RA
Ash, SR
Sutton, JM
机构
[1] HemoCleanse, W Lafayette, IN 47906 USA
[2] Arnett Clin, Lafayette, IN USA
[3] Purdue Univ, W Lafayette, IN 47907 USA
关键词
D O I
10.1097/00002480-199809000-00043
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Ash Split Cath (Medcomp, Harleysville, PA) is a recently introduced dual lumen permanent catheter designed to be placed through the internal jugular vein into the superior vena cava by single venipuncture technique. The transcutaneous portion is a 14 French cylindrically shaped catheter with D-shaped lumens and a Dacron (DuPont, Wilmington, DE) cuff. At the entrance to the jugular vein, the catheter splits into two separate D-shaped limbs that then merge into multiholed cylindrical tips in the vena cava. Split Caths (n = 10) have been placed in patients with end-stage renal disease and used for outpatient dialysis for approximately 2 months. Flow rates and hydraulic resistance have been compared with Mahurkar (Bard, Salt Lake City, UT) (n = 22) and Tesio (Medcomp) (n = 17) catheters in the same unit. Average blood flow rates (Qb) were 295 +/- 42 (SD) for Ash Split Caths vs 279 +/- 38 and 300 +/- 39 ml(3)/min for Mahurkar and Tesio catheters, respectively, and hydraulic resistances were 0.44 +/- 0.17, 0.52 +/- 0.15, and 0.56 +/- 0.11 mmHg/ml/min, respectively (not significant). No Split Caths have been removed for bleeding or flow complications. The Split Cath provides the simplicity of placement and removal of a single-bodied catheter with flow advantages of independent, cylindrical, multiholed tips.
引用
收藏
页码:M532 / M534
页数:3
相关论文
共 6 条
[1]   INTERNAL JUGULAR-VEIN CANNULATION USING 2 SILASTIC CATHETERS - A NEW, SIMPLE AND SAFE LONG-TERM VASCULAR ACCESS FOR EXTRACORPOREAL TREATMENT [J].
CANAUD, B ;
BERAUD, JJ ;
JOYEUX, H ;
MION, C .
NEPHRON, 1986, 43 (02) :133-138
[2]  
KRIVITSKI NM, 1995, ASAIO J, V41, pM741, DOI 10.1097/00002480-199507000-00111
[3]   Effective blood flow and recirculation rates in internal jugular vein twin catheters: Measurement by ultrasound velocity dilution [J].
Leblanc, M ;
Bosc, JY ;
Vaussenat, F ;
Maurice, F ;
Leray-Moragues, H ;
Canaud, B .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1998, 31 (01) :87-92
[4]   USE OF A SILICONE CATHETER WITH A DACRON CUFF FOR DIALYSIS SHORT-TERM VASCULAR ACCESS [J].
MOSS, AH ;
MCLAUGHLIN, MM ;
LEMPERT, KD ;
HOLLEY, JL .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1988, 12 (06) :492-498
[5]   SUCCESSFUL USE OF DOUBLE-LUMEN, SILICONE-RUBBER CATHETERS FOR PERMANENT HEMODIALYSIS ACCESS [J].
SHUSTERMAN, NH ;
KLOSS, K ;
MULLEN, JL .
KIDNEY INTERNATIONAL, 1989, 35 (03) :887-890
[6]   DOUBLE CATHETERIZATION OF THE INTERNAL JUGULAR-VEIN FOR HEMODIALYSIS - INDICATIONS, TECHNIQUES, AND CLINICAL-RESULTS [J].
TESIO, F ;
DEBAZ, H ;
PANARELLO, G ;
CALIANNO, G ;
QUAIA, P ;
RAIMONDI, A ;
SCHINELLA, D .
ARTIFICIAL ORGANS, 1994, 18 (04) :301-304