Performance of the VITEK2 system for identification and susceptibility testing of routine Enterobacteriaceae clinical isolates

被引:22
作者
Pérez-Vázquez, M [1 ]
Oliver, A [1 ]
del Saz, BS [1 ]
Loza, E [1 ]
Baquero, F [1 ]
Cantón, R [1 ]
机构
[1] Hosp Ramon y Cajal, Microbiol Serv, E-28034 Madrid, Spain
关键词
enterobacteriaceae; automatic susceptibility testing;
D O I
10.1016/S0924-8579(01)00318-1
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The VITEK2 system was evaluated with 138 fresh consecutive routine clinical Enterobacteriaceae isolates. Susceptibility results to 10 beta -lactams, three aminoglycosides and a quinolone were compared with those obtained following the NCCLS standard microdilution. API20E was used as reference method for identification. All but three isolates were correctly identified in 3 h at species level (97.8%), two isolates (1.4%) at genus level and only one isolate was misidentified. Overall essential agreement for susceptibility testing was 97.1%. Discrepancies were mainly observed with piperacillin (1.1%), cefuroxime (0.6%) and amoxycillin/ clavulanate (0.3%), Discrepancies for aminoglycosides and ciprofloxacin were low (< 0.1%). Minor: major and very major errors (NCCLS categories) were 4.1%, 0.2% and 6.1%, respectively. Very major errors were due to piperacillin (4.5%), ampicillin (0.8%) and amoxycillin/clavulanate (0.8%). The VITEK2 system gave accurate identification and susceptibility testing results of routine Enterobacteriaceae clinical isolates. <(c)> 2001 Published by Elsevier Science B.V. and International Society of Chemotherapy. All rights reserved.
引用
收藏
页码:371 / 376
页数:6
相关论文
共 18 条