A new high intensity focused ultrasound applicator for surgical applications

被引:33
作者
Brentnall, MD [1 ]
Martin, RW
Vaezy, S
Kaczkowski, P
Forster, F
Crum, L
机构
[1] Univ Washington, Dept Mech Engn, Seattle, WA 98195 USA
[2] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[3] Univ Washington, Dept Anesthesiol, Seattle, WA 98195 USA
[4] Univ Washington, Appl Phys Lab, Seattle, WA 98195 USA
关键词
D O I
10.1109/58.895907
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Improved high-intensity focused ultrasound (HIFU) surgical applicators are required for use in a surgical environment. We report on the performance and characteristics of a new solid-cone HIFU applicator. Previous HIFU devices used a water-filled stand-off to couple the ultrasonic energy from the transducer to the treatment area. The new applicator uses a spherically-focused element and a solid aluminum cone to guide and couple the ultrasound to the tissue. Compared with the water-filled applicators, this new applicator is simpler to set up and manipulate, cannot leak, prevents the possibility of cavitation within the coupling device, and is much easier to sterilize and maintain during surgery. The design minimizes losses caused by shear wave conversion found in tapered solid acoustic velocity transformers operated at high frequencies. Computer simulations predicted good transfer of longitudinal waves. Impedance measurements, beam plots, Schlieren images, and force balance measurements verified strong focusing and suitable transfer of acoustic energy into water. At the focus, the -3 dB beam dimensions are 1.2 mm (axial) X 0.3 mm (transverse). Radiation force balance measurements indicate a power transfer efficiency of 40%. In vitro and in vivo tissue experiments confirmed the applicator's ability to produce hemostasis.
引用
收藏
页码:53 / 63
页数:11
相关论文
共 22 条
[1]  
BALAMUTH L, 1954, T IRE PGUE 2, V1, P23
[2]  
Frederick J. R., 1965, Ultrasonic engineering
[3]  
FRY W, 1995, SCIENCE, V122, P517
[4]   PRODUCTION OF FOCAL DESTRUCTIVE LESIONS IN THE CENTRAL NERVOUS SYSTEM WITH ULTRASOUND [J].
FRY, WJ ;
MOSBERG, WH ;
BARNARD, JW ;
FRY, FJ .
JOURNAL OF NEUROSURGERY, 1954, 11 (05) :471-478
[5]  
GORDON G, 1963, MED SOC T, V29, P173
[6]  
HAMILTON M, 1998, NONLINEAR ACOUSTICS, P68
[7]   Review article: High intensity focused ultrasound-potential for cancer treatment [J].
Hill, CR ;
terHaar, GR .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (816) :1296-1303
[8]   DESIGN OF ULTRASONIC TRANSDUCERS WITH MULTIPLE ACOUSTIC MATCHING LAYERS FOR MEDICAL APPLICATION [J].
INOUE, T ;
OHTA, M ;
TAKAHASHI, S .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1987, 34 (01) :8-16
[9]  
JAMES J A, 1963, Acta Otolaryngol, V56, P148, DOI 10.3109/00016486309127399
[10]  
Kossoff G, 1979, Ultrasound Med Biol, V5, P359, DOI 10.1016/0301-5629(79)90006-1