Traceable measurements of pH

被引:30
作者
Spitzer, P [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
metrology in chemistry; traceability; pH; key comparison;
D O I
10.1007/PL00010440
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The primary method for pH is based on the measurement of the potential difference of an electrochemical cell containing a platinum hydrogen electrode and a silver/silver chloride reference electrode, often called a Harned cell. Assumptions must be made to relate the operation of this cell to the thermodynamic definition of pH. National metrology institutes use the primary method to assign pH values to a limited number of primary standards (PS). The required comparability of pH can be ensured only if the buffers used for the calibration of pH meter-electrode assemblies are traceable to there primary pH standards. To assess the degree of equivalence, comparisons of primary measurement procedures for pH were organized in co-operation with EUROMET. Typical results will be presented. In 1998, the Consultative Committee for Amount of Substance (CCQM) decided to include the field of pH in its working programme. The first key comparison for this quantity was recently carried out on two phosphate buffer solutions.
引用
收藏
页码:55 / 60
页数:6
相关论文
共 23 条
[1]  
Bates R.G., 1960, Pure Appl Chem, V1, P163, DOI [DOI 10.1351/PAC196001010163, 10.1351/pac196001010163]
[2]  
BATES RG, 1980, J SOLUTION CHEM, V9, P455
[3]  
BATES RG, 1973, DETERMINATION PH THE, P15
[4]  
BATES RG, 1973, DETERMINATION PH THE, P95
[5]   DIFFERENTIAL-POTENTIOMETRIC CELL FOR THE RESTANDARDIZATION OF PH REFERENCE MATERIALS [J].
BAUCKE, FGK .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1994, 368 (1-2) :67-75
[6]   MULTIPLE-POINT CALIBRATION WITH LINEAR-REGRESSION AS A PROPOSED STANDARDIZATION PROCEDURE FOR HIGH-PRECISION PH MEASUREMENTS [J].
BAUCKE, FGK ;
NAUMANN, R ;
ALEXANDERWEBER, C .
ANALYTICAL CHEMISTRY, 1993, 65 (22) :3244-3251
[7]  
BAUCKE FGK, 1997, W68 PTB, P10
[8]   Some recent developments in reference methodology within the United Kingdom [J].
Catterick, T ;
Craston, D ;
King, B ;
Walker, RF ;
Webb, KS .
ACCREDITATION AND QUALITY ASSURANCE, 1999, 4 (1-2) :3-13
[9]   A PITZER MIXED ELECTROLYTE SOLUTION THEORY APPROACH TO ASSIGNMENT OF PH TO STANDARD BUFFER SOLUTIONS [J].
COVINGTON, AK ;
FERRA, MIA .
JOURNAL OF SOLUTION CHEMISTRY, 1994, 23 (01) :1-10
[10]   DEFINITION OF PH SCALES, STANDARD REFERENCE VALUES, MEASUREMENT OF PH AND RELATED TERMINOLOGY - (RECOMMENDATIONS 1984) [J].
COVINGTON, AK ;
BATES, RG ;
DURST, RA .
PURE AND APPLIED CHEMISTRY, 1985, 57 (03) :531-542