Improved quantification for Tc-94m PET imaging

被引:14
作者
Barker, WC [1 ]
Szajek, LP [1 ]
Green, SL [1 ]
Carson, RE [1 ]
机构
[1] NIH, Warren Grant Magnuson Clin Ctr, PET Dept, Bethesda, MD 20892 USA
关键词
biomedical nuclear imaging; Compton scattering; poistron emission tomography (PET);
D O I
10.1109/23.940156
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The positron emitter Tc-94m (52.5-min half-life) has the potential to improve quantification of tracers currently labeled with Tc-99m. The successful production of Tc-94m and the development of an accurate multiisotope decay correction have improved quantification for Tc-94m positron emission tomography (PET) imaging. However, the presence of gamma rays emitted in coincidence with the positrons produces an apparent increase in scatter of 40% to 90%. Furthermore, the contaminant positron emitter Tc-94 (293-min half-life) produced with Tc-94m also has positron-coincident gammas, so the effect is time-dependent. In this study, modification of the kernel parameters of the Bergstrom two-dimensional (2-D) scatter correction is proposed as a simple way to improve technetium PET quantification. Parameter optimization was performed using line source scatter phantom data for F-18 and technetium for a variety of positions in cylindrical and elliptical phantoms over a 3-h period. The accuracy of the correction was evaluated with remnant scatter fraction measurements using uniformly filled phantoms with 5-cm cold cylindrical inserts. On average, remnant 2-D scatter fractions of 10 +/- 3% were observed when parameters optimized for F-18 were used. The use of parameters optimized for technetium reduced the remnant scatter fractions to 1 +/- 3%, showing that accurate quantification with Tc-94m can be achieved.
引用
收藏
页码:739 / 742
页数:4
相关论文
共 12 条
[1]   CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS [J].
BERGSTROM, M ;
ERIKSSON, L ;
BOHM, C ;
BLOMQVIST, G ;
LITTON, J .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) :42-50
[2]   IMPROVING THE RADIONUCLIDIC PURITY OF TC-94M FOR PET IMAGING [J].
CHRISTIAN, BT ;
NICKLES, RJ ;
STONE, CK ;
MULNIX, TL ;
CLARK, J .
APPLIED RADIATION AND ISOTOPES, 1995, 46 (02) :69-73
[3]  
DEGRADO TR, 1994, J NUCL MED, V35, P1398
[4]   SCATTER CORRECTION IN THE TRANSAXIAL SLICES OF A WHOLE-BODY POSITRON EMISSION TOMOGRAPH [J].
HOVERATH, H ;
KUEBLER, WK ;
OSTERTAG, HJ ;
DOLL, J ;
ZIEGLER, SI ;
KNOPP, MV ;
LORENZ, WJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) :717-728
[5]  
KOHLMYER SG, 1999, J NUCL MED, V40
[6]  
Lederer CM., 1978, Table of isotopes
[7]  
*NAT EL MAN ASSN, 1994, PERF MEAS POS EM TOM
[8]  
PENTLOW KS, 1999, J NUCL MED, V40
[9]   Production of high purity 94mTc for positron emission tomography studies [J].
Qaim, SM .
NUCLEAR MEDICINE AND BIOLOGY, 2000, 27 (04) :323-328
[10]  
ROSCH F, 1994, J NUCL MED, V35, pP253