Accuracy of density measurements within plaques located in artificial coronary arteries by X-ray multislice CT: Results of a phantom study

被引:78
作者
Schroeder, S
Flohr, T
Kopp, AF
Meisner, C
Kuettner, A
Herdeg, C
Baumbach, A
Ohnesorge, B
机构
[1] Univ Tubingen, Div Cardiol, Med Clin, Dept Internal Med, D-72074 Tubingen, Germany
[2] Univ Tubingen, Dept Radiol, Div Diagnost Radiol, D-72074 Tubingen, Germany
[3] Univ Tubingen, Inst Med Informat Proc, D-72074 Tubingen, Germany
[4] Siemens AG, Computed Tomog, Med Engn, Forchheim, Germany
关键词
coronary artery disease; computed tomography; multislice; plaques; coronary;
D O I
10.1097/00004728-200111000-00013
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Clinical studies indicate that coronary plaque morphology might be differentiated noninvasively using multislice CT by determining tissue density within the lesions. The aim of the present experimental study was to evaluate factors that influence density measurements within small vessels. Method: A coronary phantom model was developed, consisting of silicon tubes (lumen diameter 4 mm) with two plaques of known density inside, simulating soft and intermediate lesions (Plaque 1: -39 HU; Plaque 2: 72 HU). Density measurement were conducted in three different contrast medium concentrations (1:30, 1:40, 1:50) and two different slice widths (4 x 2.5 mm, 4 x 1 mm). All scans were performed on a Somatom Volume Zoom (Siemens, Forchheim, Germany). Experimental results were compared with calculated data based on computer simulation. Results: The two plaques could be clearly differentiated from each other on both collimations (4 x 2.5 mm: Plaque 1, 85 +/- 61 HU vs. Plaque 2, 91 +/- 17 HU, p < 0.0001; 4 x 1 mm: Plaque 1, 50 +/- 54 HU vs. Plaque 2. 91 +/- 17 HU, p < 0.0001). Significantly lower and more accurate results were achieved with 1.0 nim collimation (p < 0.0001). Contrast medium concentration contributed significantly to the measurements (p < 0.001). The experimental findings were confirmed by Computer simulation, which revealed even more accurate results when using a 0.5 mm collimation (Plaque 1, 0.5 mm: -9 HU vs. 4 x 1 mm: 14 HU, Plaque 2,4 x 0.5 mm: 83 HU vs. 4 x 1 mm: 93 HU). Conclusion: Density measurements were found to be highly dependent on slice width and surrounding contrast enhancement. Our results indicate that standardization of methodology is required before the noninvasive differentiation of human plaque morphology by multislice CT can be applied in the clinical setting as a screening test for coronary soft plaques.
引用
收藏
页码:900 / 906
页数:7
相关论文
共 30 条
  • [1] Lesion morphology assessed by pre-interventional intravascular ultrasound does not predict the incidence of severe coronary artery dissections
    Athanasiadis, A
    Haase, KK
    Wullen, A
    Treusch, AW
    Mahrholdt, H
    Voelker, W
    Baumbach, A
    Oberhoff, M
    Karsch, KR
    [J]. EUROPEAN HEART JOURNAL, 1998, 19 (06) : 870 - 878
  • [2] Helical and single-slice conventional CT versus electron beam CT for the quantification of coronary artery calcification
    Becker, CR
    Jakobs, TF
    Aydemir, S
    Becker, A
    Knez, A
    Schoepf, UJ
    Bruening, R
    Haberl, R
    Reiser, MF
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) : 543 - 547
  • [3] BOCKSCH WG, 1994, CORONARY ARTERY DIS, V5, P727
  • [4] Botnar RM, 2000, JMRI-J MAGN RESON IM, V11, P389, DOI 10.1002/(SICI)1522-2586(200004)11:4<389::AID-JMRI6>3.0.CO
  • [5] 2-T
  • [6] Burke AP, 2000, Z KARDIOL, V89, P49
  • [7] Magnetic resonance imaging of ischemic heart disease: Why cardiac magnetic resonance imaging will play a significant role in the management of patients with coronary artery disease
    de Roos, A
    Kunz, P
    Lamb, H
    Kroft, L
    Langerak, S
    Doornbos, J
    van der Wall, E
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1999, 23 : S135 - S141
  • [8] Coronary calcium does not accurately predict near-term future coronary events in high-risk adults
    Detrano, RC
    Wong, ND
    Doherty, TM
    Shavelle, RM
    Tang, WY
    Ginzton, LE
    Budoff, MJ
    Narahara, KA
    [J]. CIRCULATION, 1999, 99 (20) : 2633 - 2638
  • [9] Coronary calcium: The good, the bad, and the uncertain
    Doherty, TM
    Detrano, RC
    Mautner, SL
    Mautner, GC
    Shavelle, RM
    [J]. AMERICAN HEART JOURNAL, 1999, 137 (05) : 806 - 814
  • [10] Electron-beam computed tomography for detection of early signs of coronary arteriosclerosis
    Erbel, R
    Schmermund, A
    Möhlenkamp, S
    Sack, S
    Baumgart, D
    [J]. EUROPEAN HEART JOURNAL, 2000, 21 (09) : 720 - 732