Use of ion mobility spectrometry to determine low levels of impurities in gases

被引:8
作者
Ketkar, SN [1 ]
Dheandhanoo, S [1 ]
机构
[1] Air Prod & Chem Inc, Allentown, PA 18195 USA
关键词
D O I
10.1021/ac010132q
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The use of ion mobility spectrometry (IMS) for the determination of trace moisture and oxygen in bulk nitrogen has been explored. IMS utilizes atmospheric pressure ionization to ionize trace impurities in the sample gas. Mobility differences between trace impurity ions are exploited to separate these ions. Our results indicate that an IRIS can indeed be used to detect < 1 ppb O-2 and H2O in bulk nitrogen. Due to the nature of the interaction bem een trace moisture and oxygen, a multivariate calibration has to be used to obtain quantitative results, even at levels below 2 ppb.
引用
收藏
页码:2554 / 2557
页数:4
相关论文
共 9 条
[1]  
Albritton D.L., 1978, ATOM DATA NUCL DATA, V22, P1, DOI DOI 10.1016/0092-640X(78)90027-X
[2]  
EICEMAN G A, 1991, Critical Reviews in Analytical Chemistry, V22, P471, DOI 10.1080/10408349108055024
[3]   A DRIFT TUBE FOR MONITORING PPB TRACE WATER [J].
IRIE, T ;
MITSUI, Y ;
HASUMI, K .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1992, 31 (08) :2610-2615
[4]   DETERMINATION OF IMPURITIES IN GASES BY ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY [J].
KAMBARA, H ;
KANOMATA, I .
ANALYTICAL CHEMISTRY, 1977, 49 (02) :270-275
[5]   DYNAMIC DILUTION CALIBRATION SYSTEM FOR CALIBRATING ANALYTICAL INSTRUMENTS USED IN GAS-ANALYSIS [J].
KETKAR, SN ;
SCOTT, AD ;
DEPINILLOS, JVM .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1994, 141 (01) :184-187
[6]  
KETKAR SN, 1994, MICROCONTAMINATI JAN
[7]  
*PCP INC, MOD 111 PCP
[8]  
SIEFERING K, 1994, J VAC SCI TECHNOL A, V11, P4
[9]  
STIMAC R, COMMUNICATION