Device-based change in left ventricular shape: A new concept for the treatment of dilated cardiomyopathy

被引:68
作者
McCarthy, PM
Takagaki, M
Ochiai, Y
Young, JB
Tabata, T
Shiota, T
Qin, JX
Thomas, JD
Mortier, TJ
Schroeder, RF
Schweich, CJ
Fukamachi, K
机构
[1] Cleveland Clin Fdn, Dept Thorac & Cardiovasc Surg, Kaufman Ctr Heart Failure, Cleveland, OH 44195 USA
[2] Cleveland Clin Fdn, Dept Cardiol, Kaufman Ctr Heart Failure, Cleveland, OH 44195 USA
[3] Cleveland Clin Fdn, Lerner Res Inst, Dept Biomed Engn, Cleveland, OH 44195 USA
[4] Myocor Inc, Maple Grove, MN USA
关键词
D O I
10.1067/mtc.2001.115240
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: We tested a unique new device, the Myosplint device (Myocor, Inc, Maple Grove, Minn), which is designed to change left ventricular shape, reduce left ventricular wall stress, and improve left ventricular systolic function. Methods: Heart failure was induced in 15 dogs over 27 days by rapid pacing (230 beats/min). Seven animals underwent sham surgery, and 8 animals received 3 transventricular Myosplint devices each. Myosplint devices were tightened to create a symmetric bilobular left ventricular shape and were adjusted to produce a calculated 20% reduction in wall stress. Hemodynamic, 2-dimensional, and 3-dimensional echocardiographic studies were recorded at baseline, immediately after Myosplint placement (acute change), and at 1 month after both groups had a reduced rate (190 beats/min) of pacing designed to maintain heart failure. Results: The Myosplint group had significant sustained improvements in left ventricular ejection fraction from baseline, to the acute change, to 1 month (19% +/- 5%; 36% +/- 8% ; 39% +/- 13%) and reductions of left ventricular end-systolic volumes (73 +/- 9 mL; 34 +/- 5 mL; 42 +/- 12 mL) and end-systolic wall stress by 39% (341 +/- 68 10(3) dynes . cm(-2) to 206 +/- 28 10(3) dynes . cm(-2)) acutely and 31% (372 +/- 83 10(3) dynes . cm(-2) to 250 +/- 40 10(3) dynes . cm(-2)) at 1 month. There were no significant changes in mitral regurgitation. Conclusion: Application of a Myosplint device to a dilated impaired left ventricle resulted in reduced wall stress and improved left ventricular systolic function that was sustained at 1 month. Device-based shape change is a promising new opportunity to treat patients with dilated cardiomyopathy.
引用
收藏
页码:482 / 490
页数:9
相关论文
共 25 条
[1]  
ARMSTRONG PW, 1996, PATHOPHYSIOLOGY TACH, P45
[2]   Partial left ventriculectomy to improve left ventricular function in end-stage heart disease [J].
Batista, RJV ;
Santos, JLV ;
Takeshita, N ;
Bocchino, L ;
Lima, PN ;
Cunha, MA .
JOURNAL OF CARDIAC SURGERY, 1996, 11 (02) :96-97
[3]   Heart reduction surgery: An analysis of the impact on cardiac function [J].
Dickstein, ML ;
Spotnitz, HM ;
Rose, EA ;
Burkhoff, D .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1997, 113 (06) :1032-1040
[4]  
DIDONATO M, 1995, EUR HEART J, V16, P1285
[5]  
FRANCIS GS, 1999, TXB CLIN DISORDERS A, P477
[6]   Partial left ventriculectomy for dilated cardiomyopathy: Is this an alternative to transplantation? [J].
Franco-Cereceda, A ;
McCarthy, PM ;
Blackstone, EH ;
Hoercher, KJ ;
White, JA ;
Young, JB ;
Starling, RC .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 121 (05) :879-893
[7]   SYSTOLIC AND DIASTOLIC DYSFUNCTION DURING ATRIAL-PACING IN CONSCIOUS DOGS WITH LEFT-VENTRICULAR HYPERTROPHY [J].
FUJII, AM ;
GELPI, RJ ;
MIRSKY, I ;
VATNER, SF .
CIRCULATION RESEARCH, 1988, 62 (03) :462-470
[8]  
Fukamachi K., 2000, Journal of Heart and Lung Transplantation, V19, P68
[9]  
GREENBAUM RA, 1981, BRIT HEART J, V45, P248
[10]   ISOPROTERENOL-INDUCED ALTERATIONS IN MYOCARDIAL BLOOD-FLOW, SYSTOLIC AND DIASTOLIC FUNCTION IN CONSCIOUS DOGS WITH HEART-FAILURE [J].
HITTINGER, L ;
SHANNON, RP ;
KOHIN, S ;
LADER, AS ;
MANDERS, WT ;
PATRICK, TA ;
KELLY, P ;
VATNER, SF .
CIRCULATION, 1989, 80 (03) :658-668