Haemodialysing infants: theoretical limitations, and single versus double lumen lines

被引:12
作者
Coulthard, MG [1 ]
Sharp, J [1 ]
机构
[1] Royal Victoria Infirm, Dept Paediat Nephrol, Newcastle Upon Tyne NE1 4LP, Tyne & Wear, England
关键词
haemodialysis; infants; vascular access; blood flow; dialysis adequacy;
D O I
10.1007/s004670000547
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Haemodialysing small infants is difficult because of vascular access limitations. We show that Poiseuille's law (that now through a tube varies with its radius(4)) makes it inevitable that the blood flow that can be achieved in smaller patients will fall disproportionately compared to their need for dialysis. Poiseuille's law also predicts that for single and multiple lumen cannulae of the same outside gauge, blood flow through the single lu men will be several times greater. Measurements confirmed this. It is argued that haemodialysis efficiency will therefore be improved by using a single lumen cannula to alternately withdraw and return blood, compared to sampling and returning continuously through a multiple lumen cannula, despite only withdrawing for half the time.
引用
收藏
页码:332 / 334
页数:3
相关论文
共 4 条
[1]  
[Anonymous], 1935, The growth of the surface area of the human body
[2]   WEIGHT AS THE BEST STANDARD FOR GLOMERULAR-FILTRATION IN THE NEWBORN [J].
COULTHARD, MG ;
HEY, EN .
ARCHIVES OF DISEASE IN CHILDHOOD, 1984, 59 (04) :373-375
[3]   HEMODIALYSIS AND ULTRAFILTRATION IN BABIES WEIGHING UNDER 1000 G [J].
COULTHARD, MG ;
SHARP, J .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1995, 73 (03) :F162-F165
[4]   Dialysis in neonates with inborn errors of metabolism [J].
Schaefer, F ;
Straube, E ;
Oh, J ;
Mehls, O ;
Mayatepek, E .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1999, 14 (04) :910-918