Miniaturized implantable rotary blood pump in atrial-aortic position supports and unloads the failing heart

被引:20
作者
Meyns, B
Siess, T
Nishimura, Y
Racz, R
Reul, H
Rau, G
Leunens, V
Flameng, W
机构
[1] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, D-5100 Aachen, Germany
[2] Katholieke Univ Leuven, Dept Cardiac Surg, Louvain, Belgium
来源
CARDIOVASCULAR SURGERY | 1998年 / 6卷 / 03期
关键词
assist device; rotary blood pump; heart failure;
D O I
10.1016/S0967-2109(97)00170-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to test the pump performance of a miniaturized, newly developed, implantable rotary blood pump (the diagonal pump). In a series of six sheep, the pump was interposed in an atrial-aortic connection. Heart failure was induced by serial injection of glass beads into the left coronary artery, The assistance of the non-failing heart (with pump output up to 4.5 liter/min) did not change the total cardiac output or the blood pressure of the animal. The heart was significantly unloaded as demonstrated by a drop in first derivative of the left ventricular pressure (dP/dt) max (from 1645 to 1113 mmHg/s; P = 0.0003). Because of the specific interaction between heart function and rotary blood pump performance, the pump provided considerably more flow in the failing than in the non-failing heart. Cardiac output and perfusion pressures were restored to pre-assist values in the failing heart model. The heart-rotary blood pump interaction is an ideal substrate for long-term assistance. With this miniaturized rotary blood pump, hemodynamics in a severely failing heart can be restored. Atrial cannulation, which leaves the left ventricle untouched, provides similar results as apical cannulation in the failing heart. (C) 1998 The International Society for Cardiovascular Surgery. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:288 / 295
页数:8
相关论文
共 10 条
[1]  
GROSS DR, 1994, ANIMAL MODELS CARDIO, P421
[2]   POTENTIAL OF LEFT-VENTRICULAR ASSIST DEVICES AS OUTPATIENT THERAPY WHILE AWAITING TRANSPLANTATION [J].
LEVIN, HR ;
CHEN, JM ;
OZ, MC ;
CATANESE, KA ;
KRUM, H ;
GOLDSMITH, RL ;
PACKER, M ;
ROSE, EA .
ANNALS OF THORACIC SURGERY, 1994, 58 (05) :1515-1520
[3]  
MCCARTHY PM, 1991, J THORAC CARDIOV SUR, V102, P578
[4]  
MEHTA SM, 1995, J HEART LUNG TRANSPL, V14, P585
[5]  
MEYNS B, 1995, INT J ARTIF ORGANS, V18, P452
[6]   LEFT-VENTRICULAR ASSISTANCE WITH THE TRANSTHORACIC 24F HEMOPUMP FOR RECOVERY OF THE FAILING HEART [J].
MEYNS, BP ;
SERGEANT, PT ;
DAENEN, WJ ;
FLAMENG, WJ .
ANNALS OF THORACIC SURGERY, 1995, 60 (02) :392-397
[7]  
MUELLER J, 1995, INT J ARTIF ORGANS, V18, P453
[8]  
PAE WE, 1992, J THORAC CARDIOV SUR, V104, P541
[9]   TECHNICAL REQUIREMENTS AND LIMITATIONS OF MINIATURIZED AXIAL-FLOW PUMPS FOR CIRCULATORY SUPPORT [J].
REUL, H .
CARDIOLOGY, 1994, 84 (03) :187-193
[10]   Hemodynamic system analysis of intraarterial microaxial pumps in vitro and in vivo [J].
Siess, T ;
Meyns, B ;
Spielvogel, K ;
Reul, H ;
Rau, G ;
Flameng, W .
ARTIFICIAL ORGANS, 1996, 20 (06) :650-661