Ultratrace determination of cadmium by atomic absorption spectrometry using hydride generation with in situ preconcentration in a palladium-coated graphite atomizer

被引:44
作者
Infante, HG [1 ]
Sanchez, MLF [1 ]
SanzMedel, A [1 ]
机构
[1] UNIV OVIEDO,DEPT PHYS & ANALYT CHEM,OVIEDO 33006,SPAIN
关键词
cadmium hydride generation; in situ preconcentration; palladium coated graphite tube; electrothermal atomic absorption spectrometry;
D O I
10.1039/ja9961100571
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A very sensitive ETAAS method, with in situ trapping of the metal, is described for ultratrace determinations of Cd in liquid solutions. It is based on volatile Cd HG from a vesicular solution with subsequent trapping of the metal in a graphite platform coated with Pd. This Pd coating, on a pyrolytic graphite L'vov platform, acts as the atomizer, Sodium tetrahydroborate(nr) in a vesicular medium of didodecyldimethylammonium bromide is used to generate Cd hydride from the sample solution, This volatile species is trapped by the Pd coating, pre-heated at 150 degrees C, the preconcentrated Cd then atomized at 1600 degrees C and the transient AAS signal measured, Detection limits (DL) observed are in the ng l(-1) range (e.g., using a 1.4 ml Cd sample volume for preconcentration the DL was 60 ng l(-1)). Higher preconcentrations up to a maximum of 1.4 ng of total Cd deposited are possible with the proposed methodology, The RSD for ten replicate analyses of 1.4 mi of 500 ng l(-1) Cd solution (0.7 ng of total Cd) was 1.7%. The combination of vesicles-mediated HG with the ease of decomposition of Cd hydride on the Pd coating at 150 degrees C, provides a convenient Cd preconcentration method for enhancing the sensitivity of direct determinations of Cd by conventional ETAAS or by HGAAS.
引用
收藏
页码:571 / 575
页数:5
相关论文
共 26 条
[1]  
CHAUDHRY MM, 1991, ANAL P, V28, P45
[2]  
DEDINA J, 1995, HYDRIDE GENERATION A
[3]  
DOIDGE PS, 1989, J ANAL ATOM SPECTROM, V4, P251, DOI 10.1039/ja9890400251
[4]  
Fernandez De La Campa M. R., 1995, Spectrochimica Acta Part B Atomic Spectroscopy, V50, P377, DOI 10.1016/0584-8547(94)00155-O
[5]   GENERATION OF COVALENT HYDRIDES IN ATOMIC-ABSORPTION SPECTROSCOPY - A REVIEW [J].
GODDEN, RG ;
THOMERSON, DR .
ANALYST, 1980, 105 (1257) :1137-1156
[6]   GAS-SAMPLING TECHNIQUE FOR ARSENIC DETERMINATION BY ATOMIC ABSORPTION SPECTROPHOTOMETRY [J].
HOLAK, W .
ANALYTICAL CHEMISTRY, 1969, 41 (12) :1712-&
[8]   INDIUM HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY WITH IN-SITU PRECONCENTRATION IN A GRAPHITE-FURNACE COATED WITH PALLADIUM [J].
LIAO, YP ;
LI, AM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (04) :633-636
[9]   DETERMINATION OF SELENIUM IN MARINE ORGANISMS USING HYDRIDE GENERATION AND ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROSCOPY [J].
MAHER, WA .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1983, 16 (11) :801-810
[10]  
NAKAHARA T, 1983, PROG ANAL ATOM SPECT, V6, P163