Characterizing output for dynamic wedges

被引:16
作者
Liu, CR
Zhu, TC
Palta, JR
机构
[1] Department of Radiation Oncology, College of Medicine, University of Florida, Gainesville
关键词
dynamic wedge; segmented treatment tables; output factor;
D O I
10.1118/1.597677
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The output factor for the dynamic wedge, unlike that for the physical wedge, is a complex function of the field dimension along the moving jaw and wedge angle. The large change in output (varying as much as 40% for 45 degrees and 60 degrees wedge angles) can be attributed clearly to the segmented treatment tables (STTs), which specify cumulative monitor unit weighting as a function of jaw position, y. We found that the output factor (in air or water) on the central axis for the dynamic wedge can be characterized by multiplying the output factor (in air or water) for an open field by a normalization factor, which is determined from the STTs, thus indicating that collimator scatter is similar for both the dynamic wedge field and the open field. The introduction of the normalization factor decreases the commissioning: time for dynamic wedges significantly and is useful for quality assurance. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:1213 / 1218
页数:6
相关论文
共 8 条
[1]   THE RELATION BETWEEN WEDGE FACTORS IN AIR AND WATER [J].
BARDEROMA, RD ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1994, 21 (07) :1043-1047
[2]   BEAM DATA MEASUREMENTS FOR DYNAMIC WEDGES ON VARIAN-600C (6 MV) AND VARIAN-2100C (6 AND 10 MV) LINEAR ACCELERATORS [J].
BIDMEAD, AM ;
GARTON, AJ ;
CHILDS, PJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (03) :393-411
[3]  
KIJEWSKI P K, 1978, Medical Physics (Woodbury), V5, P426, DOI 10.1118/1.594440
[4]   DOSIMETRY AND CLINICAL IMPLEMENTATION OF DYNAMIC WEDGE [J].
KLEIN, EE ;
LOW, DA ;
MEIGOONI, AS ;
PURDY, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (03) :583-592
[5]   DYNAMIC WEDGE FIELD TECHNIQUES THROUGH COMPUTER-CONTROLLED COLLIMATOR MOTION AND DOSE DELIVERY [J].
LEAVITT, DD ;
MARTIN, M ;
MOELLER, JH ;
LEE, WL .
MEDICAL PHYSICS, 1990, 17 (01) :87-91
[6]  
LEAVITT DD, 1992, MED DOSIMETRY, V15, P47
[7]   THE DETERMINATION OF PHANTOM AND COLLIMATOR SCATTER COMPONENTS OF THE OUTPUT OF MEGAVOLTAGE PHOTON BEAMS - MEASUREMENT OF THE COLLIMATOR SCATTER PART WITH A BEAM-COAXIAL NARROW CYLINDRICAL PHANTOM [J].
VANGASTEREN, JJM ;
HEUKELOM, S ;
VANKLEFFENS, HJ ;
VANDERLAARSE, R ;
VENSELAAR, JLM ;
WESTERMANN, CF .
RADIOTHERAPY AND ONCOLOGY, 1991, 20 (04) :250-257
[8]  
WALDRON TJ, 1994, MED PHYS, V21, P873