Modular tibial implants submitted for retrieval analysis were examined for evidence of cold flow, wear, or polyethylene failure. All retrieved components showed areas of cold flow and wear. Significant damage, defined as pitting, gouging, or delamination was observed in 77% of the retrieved implants. Cold flow deformation commonly was observed at the junction of the polyethylene and locking mechanisms or junction of the polyethylene and the edge of the tibial tray. Control implants did not show such areas of damage. Clinical factors such as age, activity status, length of implantation, limb alignment before revision, and thickness of the tibial insert did not correlate with backside wear. No association was found between polyethylene thickness, time of implantation, limb or prosthesis alignment, or activity status or weight of the patient.
引用
收藏
页码:248 / 253
页数:6
相关论文
共 11 条
[1]  
APEL DM, 1991, CLIN ORTHOP RELAT R, P243
[2]   PERFORMANCE OF THE TIBIAL COMPONENT IN TOTAL KNEE REPLACEMENT - CONVENTIONAL AND REVISION DESIGNS [J].
BARTEL, DL ;
BURSTEIN, AH ;
SANTAVICCA, EA ;
INSALL, JN .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (07) :1026-1033
[3]  
CALLAGHAN JJ, 1995, CLIN ORTHOP RELAT R, P76
[4]  
Font-Rodriguez DE, 1997, CLIN ORTHOP RELAT R, P79
[5]   RESULTS OF LOW FRICTION ARTHROPLASTY IN PATIENTS 30 YEARS OF AGE OR YOUNGER [J].
HALLEY, DK ;
CHARNLEY, J .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1975, (112) :180-191
[6]  
HUNGERFORD DS, 1983, CLIN ORTHOP RELAT R, P95
[7]   DISTINGUISHING WEAR AND CREEP IN CLINICALLY RETRIEVED POLYETHYLENE INSERTS [J].
MCDONALD, MD ;
BLOEBAUM, RD .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (01) :1-7
[8]  
RITTER MA, 1994, ORTHOPEDICS, V17, P841
[9]  
SALVATI EA, 1995, CLIN ORTHOP RELAT R, P85
[10]   The role of access of joint fluid to bone in periarticular osteolysis - A report of four cases [J].
Schmalzried, TP ;
Akizuki, KH ;
Fedenko, AN ;
Mirra, J .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (03) :447-452