The aim of this paper is to examine signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and image quality of cardiac CINE imaging at 1.5 T and 3.0 T. Twenty volunteers underwent cardiac magnetic resonance imaging (MRI) examinations using a 1.5-T and a 3.0-T scanner. Three different sets of breath-held, electrocardiogram-gated (ECG) CINE imaging techniques were employed, including: (1) unaccelerated SSFP (steady state free precession), (2) accelerated SSFP imaging and (3) gradient-echo-based myocardial tagging. Two-dimensional CINE SSFP at 3.0 T revealed an SNR improvement of 103% and a CNR increase of 19% as compared to the results obtained at 1.5 T. The SNR reduction in accelerated 2D CINE SSFP imaging was larger at 1.5 T (37%) compared to 3.0 T (26%). The mean SNR and CNR increase at 3.0 T obtained for the tagging sequence was 88% and 187%, respectively. At 3.0 T, the duration of the saturation bands persisted throughout the entire cardiac cycle. For comparison, the saturation bands were significantly diminished at 1.5 T during end-diastole. For 2D CINE SSFP imaging, no significant difference in the left ventricular volumetry and in the overall image quality was obtained. For myocardial tagging, image quality was significantly improved at 3.0 T. The SNR reduction in accelerated SSFP imaging was overcompensated by the increase in the baseline SNR at 3.0 T and did not result in any image quality degradation. For cardiac tagging techniques, 3.0 T was highly beneficial, which holds the promise to improve its diagnostic value.
引用
收藏
页码:1586 / 1597
页数:12
相关论文
共 35 条
[1]   Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences [J].
Alfakih, K ;
Plein, S ;
Thiele, H ;
Jones, T ;
Ridgway, JP ;
Sivananthan, MU .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (03) :323-329
[2]  
[Anonymous], P 42 ANN AD ED RES C
[3]   TAGASIST: A post-processing and analysis tools package for tagged Magnetic Resonance Imaging [J].
Bundy, JM ;
Lorenz, CH .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1997, 21 (04) :225-232
[4]  
Friedrich MG, 2000, MAGN RESON MATER PHY, V11, P16, DOI 10.1016/S1352-8661(00)00101-0
[5]   Double inversion black-blood fast spin-echo imaging of the human heart: A comparison between 1.5T and 3.0T [J].
Greenman, RL ;
Shirosky, JE ;
Mulkern, RV ;
Rofsky, NM .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (06) :648-655
[6]   Comparison of different cardiac MRI sequences at 1.5 T/3.0 T with respect to signal-to-noise and contrast-to-noise ratios - Initial experience [J].
Gutberlet, M ;
Spors, B ;
Grothoff, M ;
Freyhardt, P ;
Schwinge, K ;
Plotkin, M ;
Amthauer, H ;
Noeske, R ;
Felix, R .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (06) :801-808
[7]  
Gutberlet M, 2003, ROFO-FORTSCHR RONTG, V175, P942
[8]  
GUTBERLET M, 2004, IMAGING DECIS, V8, P23
[9]   Comparison of cardiac MRI on 1.5 and 3.0 Tesla clinical whole body systems [J].
Hinton, DP ;
Wald, LL ;
Pitts, J ;
Schmitt, F .
INVESTIGATIVE RADIOLOGY, 2003, 38 (07) :436-442
[10]   T2weighted half-Fourier acquired single-shot turbo-spin-echo-sequence in comparison with standard T2-weighted fast-spin-echo-sequence for cerebral MRI.: A sequence comparison. [J].
Hoffmann, KT ;
Hosten, N ;
Ehrenstein, T ;
Gutberlet, M ;
Röricht, S ;
Felix, R .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (06) :521-526