Neonatal infrared axillary thermometry

被引:7
作者
Seguin, J [1 ]
Terry, K [1 ]
机构
[1] Ohio State Univ, Dept Pediat, Columbus, OH 43210 USA
关键词
D O I
10.1177/000992289903800105
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The authors compared axillary skin temperatures (AT) measured with an infrared (IR) thermometer (Lightouch Neonate, Exergen Corp) with rectal temperatures (RT) in 16 newly born term infants under radiant warmers (RW) and in cribs. Twelve stable, growing premature infants in incubators were also studied. This new device may be useful because of safety and rapid results (1 second), but clinical accuracy is unknown. For term infants, mean (SD) RT-AT difference was 0.1 (0.48)degrees C under RW and 0.25 (0.17)degrees C 2 hours later in cribs, For premature infants in incubators the mean RT-AT difference was 0.09 (0.16)degrees C. Axillary temperatures measured by IR thermometer approximate RT for newly born term infants in cribs and stable premature infants in incubators, For newly born term infants under RW, RT-AT differences vary more widely, limiting clinical usefulness ia this setting. The device, the unique age of this population, and the RW environment may play a role.
引用
收藏
页码:35 / 40
页数:6
相关论文
共 8 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]  
EOFF MJF, 1974, NURS RES, V23, P457
[3]  
HADDOCK BJ, 1988, NEONATAL NETWORK, V7, P67
[4]  
JOHNSON KJ, 1991, PEDIATRICS, V87, P34
[5]   TEMPERATURE-MEASUREMENT IN TERM AND PRETERM NEONATES [J].
MAYFIELD, SR ;
BHATIA, J ;
NAKAMURA, KT ;
RIOS, GR ;
BELL, EF .
JOURNAL OF PEDIATRICS, 1984, 104 (02) :271-275
[6]  
SCHIFFMAN RF, 1982, NURS RES, V31, P274
[7]  
Weiss M E, 1994, Neonatal Netw, V13, P35
[8]   COMPARISON OF TEMPERATURE-MEASUREMENTS BY AN AURAL INFRARED THERMOMETER WITH MEASUREMENTS BY TRADITIONAL RECTAL AND AXILLARY TECHNIQUES [J].
YETMAN, RJ ;
COODY, DK ;
WEST, MS ;
MONTGOMERY, D ;
BROWN, M .
JOURNAL OF PEDIATRICS, 1993, 122 (05) :769-773