Design and performance of a low-flow capillary electrophoresis-electrospray-mass spectrometry interface using an emitter with dual beveled edge

被引:25
作者
Chen, YR [1 ]
Tseng, MC [1 ]
Her, GR [1 ]
机构
[1] Natl Taiwan Univ, Dept Chem, Taipei 10764, Taiwan
关键词
capillary zone electrophoresis; ganglioside; low-flow interface; mass spectrometry; nonvolatile buffer;
D O I
10.1002/elps.200410159
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A low-flow electrophoresis-mass spectrometry (CE-MS) interface has been developed for interfacing capillary zone electrophoresis (CZE) with electrospray- ionization-mass spectrometry (ESI-MS). The interface consists of two parallel capillary columns (a separation column and a makeup column), and an emitter with a dual beveled edge. While maintaining a relatively low optimum flow rate, the dual-beveled-edge ESI emitter allows the use of a tip with larger orifice. Therefore, this interface is less prone to column blocking in comparison with a flat tip. Primarily attributed to low sample dilution and smaller initial droplet, the interface showed better sensitivity than a conventional sheath liquid interface. Furthermore, the interface was found to be more resistant to the presence of nonvolatile salts. By using 40 mm borate and 20 mm alpha-cyclodextrin (alpha-CD) as the running buffer, four major forms of gangliosides were detected by CE-MS.
引用
收藏
页码:1376 / 1382
页数:7
相关论文
共 36 条
[1]  
Abian J, 1999, J MASS SPECTROM, V34, P244, DOI 10.1002/(SICI)1096-9888(199904)34:4<244::AID-JMS775>3.0.CO
[2]  
2-0
[3]  
ALEKSANDROV ML, 1984, BIOORG KHIM+, V10, P710
[4]  
ALEXANDROV ML, 1984, DOKL AKAD NAUK SSSR+, V277, P379
[5]  
Barroso MB, 1999, J AM SOC MASS SPECTR, V10, P1271
[6]   CAPILLARY ELECTROPHORESIS MASS-SPECTROMETRY [J].
CAI, JY ;
HENION, J .
JOURNAL OF CHROMATOGRAPHY A, 1995, 703 (1-2) :667-692
[7]   Sheathless capillary electrophoresis/electrospray mass spectrometry using a carbon-coated tapered fused-silica capillary with a beveled edge [J].
Chang, YZ ;
Chen, YR ;
Her, GR .
ANALYTICAL CHEMISTRY, 2001, 73 (21) :5083-5087
[8]   Sheathless capillary electrophoresis/electrospray mass spectrometry using a carbon-coated fused silica capillary [J].
Chang, YZ ;
Her, GR .
ANALYTICAL CHEMISTRY, 2000, 72 (03) :626-630
[9]   A simple method for fabrication of silver-coated sheathless electrospray emitters [J].
Chen, YR ;
Her, GR .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (05) :437-441
[10]   A low-flow CE/electrospray ionization MS interface for capillary zone electrophoresis, large-volume sample stacking, and micellar electrokinetic chromatography [J].
Chen, YR ;
Tseng, MC ;
Chang, YZ ;
Her, GR .
ANALYTICAL CHEMISTRY, 2003, 75 (03) :503-508