Selective dysfunction of ventricular electrode-endocardial junction following DC cardioversion in a patient with a dual chamber pacemaker

被引:11
作者
Das, G
Staffanson, DB
机构
[1] VET ADM MED CTR,LAS VEGAS,NV
[2] MEDTRONIC INC,MINNEAPOLIS,MN
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 02期
关键词
pacemakers; pacing threshold; thermal burn; direct current cardioversion; pacemaker malfunction;
D O I
10.1111/j.1540-8159.1997.tb06184.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A patient with a dual chamber pacemaker received two transthoracic (anterior-anterior paddle position) shocks for VF. Subsequent analysis of stimulation thresholds revealed a marked rise in the ventricular threshold only. Later, she received three additional shocks (anterior-posterior paddle position) during electrophysiological study to terminate episodes of induced VT. Following these, the ventricular stimulation threshold was > 10 V at 2.0 ms. The selective damage at the ventricular site in this patient led us to carry out in vitro studies that revealed a preferential shunting of high electrical energy into the ventricular lead, as compared to the atrial lead, following a DC shock. These observations may explain the selective pacing malfunction observed in our patient.