Lowering the detection limit of solvent polymeric ion-selective membrane electrodes. 2. Influence of composition of sample and internal electrolyte solution

被引:274
作者
Sokalski, T
Ceresa, A
Fibbioli, M
Zwickl, T
Bakker, E
Pretsch, E
机构
[1] ETH Zurich, Dept Organ Chem, CH-8092 Zurich, Switzerland
[2] Auburn Univ, Dept Chem, Auburn, AL 36849 USA
关键词
D O I
10.1021/ac9809332
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The influence of the composition of the internal electrolyte solution on the response of Pb2+- and Ca2+-selective membrane electrodes is investigated. It is shown that the lower detection limit is improved by generating, in the membrane, ionic gradients that lead to a flux of primary ions toward the inner reference electrolyte solution. If the ion flux is too strong, it may cause analyte depletion at the membrane surface and, as a consequence, apparent super-Nernstian response. Such electrodes are not adequate to measure low analyte activities but can be used to determine unbiased selectivity factors. The results are interpreted in terms of a steady-state model, introduced in the companion paper, that describes the influence of concentration gradients generated by ion-exchange and coextraction processes on both sides of the membrane.
引用
收藏
页码:1210 / 1214
页数:5