Capillary LC-MS2 at the attomole level for monitoring and discovering endogenous peptides in microdialysis samples collected in vivo

被引:110
作者
Haskins, WE [1 ]
Wang, ZQ [1 ]
Watson, CJ [1 ]
Rostand, RR [1 ]
Witowski, SR [1 ]
Powell, DH [1 ]
Kennedy, RT [1 ]
机构
[1] Univ Florida, Dept Chem, Gainesville, FL 32611 USA
关键词
D O I
10.1021/ac010774d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Fused-silica capillary LC columns (25-mum i.d.) with 3-mum-i.d. integrated electrospray emitters interfaced to a quadrupole ion trap mass spectrometer were evaluated for high-sensitivity LC-MS2. Column preparation involved constructing frits by in situ photopolymerization of glycidyl methacrylate and trimethylolpropane trimethacrylate, preparing the electrospray emitter by pulling the column outlet to a fine tip with a CO2 laser puller, and slurry-packing the column with 5-mum reversed-phase particles. Large-volume injections were facilitated by an automated two-pump system that allowed high-flow rates for sample loading and low-flow rates for elution. Small electrospray emitters, low elution flow rates, and optimization of gradient steepness allowed a detection limit of 4 amol, corresponding to 2 pM for 1.8 muL injected on-column, for a mixture of peptides dissolved in artificial cerebral spinal fluid. The system was coupled on-line to microdialysis sampling and was used to monitor and discover endogenous neuropeptides from the globus pallidus of anesthetized. male Sprague-Dawley rats. Time-segmented MS2 scans enabled simultaneous monitoring of Met-enkephalin, Leu-enkephalin, and unknown peptides. Basal dialysate levels of Met-enkephalin and Leu-enkephalin were 60 +/- 30 and 70 +/- 20 pM while K-(+)-stimulated levels were 1900 +/- 500 and 1300 +/- 300 pM, respectively (n = 7). Data-dependent and time-segmented MS2 scans revealed several unknown peptides that were present in dialysate. One of the unknowns was identified as peptide I1-10 (SPQLEDEAKE), a novel product of preproenkephalin A processing, using MS2, MS3, and database searching.
引用
收藏
页码:5005 / 5014
页数:10
相关论文
共 29 条
[1]   Determination of extracellular release of neurotensin in discrete rat brain regions utilizing in vivo microdialysis/electrospray mass spectrometry [J].
Andrén, PE ;
Caprioli, RM .
BRAIN RESEARCH, 1999, 845 (02) :123-129
[2]  
BAIRD A, 1984, P SOC EXP BIOL MED, V175, P304
[3]   Macroporous photopolymer frits for capillary electrochromatography [J].
Chen, JR ;
Dulay, MT ;
Zare, RN ;
Svec, F ;
Peters, E .
ANALYTICAL CHEMISTRY, 2000, 72 (06) :1224-1227
[4]   The minor histocompatibility antigen HA-1: A diallelic gene with a single amino acid polymorphism [J].
den Haan, JMM ;
Meadows, LM ;
Wang, W ;
Pool, J ;
Blokland, E ;
Bishop, TL ;
Reinhardus, C ;
Shabanowitz, J ;
Offringa, R ;
Hunt, DF ;
Engelhard, VH ;
Goulmy, E .
SCIENCE, 1998, 279 (5353) :1054-1057
[5]   Specific molecular mass detection of endogenously released neuropeptides using in vivo microdialysis mass spectrometry [J].
Emmett, MR ;
Andren, PE ;
Caprioli, RM .
JOURNAL OF NEUROSCIENCE METHODS, 1995, 62 (1-2) :141-147
[6]   Processing of proenkephalin-A in bovine chromaffin cells - Identification of natural derived fragments by N-terminal sequencing and matrix-assisted laser desorption ionization-time of flight mass spectrometry [J].
Goumon, Y ;
Lugardon, K ;
Gadroy, P ;
Strub, JM ;
Welters, ID ;
Stefano, GB ;
Aunis, D ;
Metz-Boutigue, MH .
JOURNAL OF BIOLOGICAL CHEMISTRY, 2000, 275 (49) :38355-38362
[7]   DIRECT IDENTIFICATION OF AN ENDOGENOUS PEPTIDE RECOGNIZED BY MULTIPLE HLA-A2.1-SPECIFIC CYTOTOXIC T-CELLS [J].
HENDERSON, RA ;
COX, AL ;
SAKAGUCHI, K ;
APPELLA, E ;
SHABANOWITZ, J ;
HUNT, DF ;
ENGELHARD, VH .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1993, 90 (21) :10275-10279
[8]   Characterization of peptides bound to the class I MHC molecule HLA-A2.1 by mass spectrometry (Reprinted from Science, vol 255, March 6, 1992) [J].
Hunt, Donald F. ;
Henderson, Robert A. ;
Shabanowitz, Jeffrey ;
Sakaguchi, Kazuyasu ;
Michel, Hanspeter ;
Sevilir, Noelle ;
Cox, Andrea L. ;
Appella, Ettore ;
Engelhard, Victor H. .
JOURNAL OF IMMUNOLOGY, 2007, 179 (05) :2669-2671
[9]   PREPARATION AND EVALUATION OF PACKED CAPILLARY LIQUID-CHROMATOGRAPHY COLUMNS WITH INNER DIAMETERS FROM 20-MU-M TO 50-MU-M [J].
KENNEDY, RT ;
JORGENSON, JW .
ANALYTICAL CHEMISTRY, 1989, 61 (10) :1128-1135
[10]   BRAIN PEPTIDES - WHAT, WHERE, AND WHY [J].
KRIEGER, DT .
SCIENCE, 1983, 222 (4627) :975-985