Simultaneous macro and micro dosimetry with MOSFETs

被引:35
作者
Rosenfeld, AB
Kaplan, GI
Carolan, MG
Allen, BJ
Maughan, R
Yudelev, M
Kota, C
Coderre, J
机构
[1] ST GEORGE HOSP,CANC CARE CTR,KOGARAH,NSW 2217,AUSTRALIA
[2] ILLAWARRA CANC CARE CTR,WOLLONGONG,NSW 2500,AUSTRALIA
[3] WAYNE STATE UNIV,HARPER HOSP,DETROIT,MI 48201
[4] BROOKHAVEN NATL LAB,UPTON,NY 11973
关键词
D O I
10.1109/23.556855
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The application of MOSFET dosimeters in complicated mixed radiation fields for measurement of absorbed dose distribution in tissue equivalent phantoms has been studied. The spectra of secondary charged particles have been measured simultaneously with average absorbed doses by the same MOSFET dosimeter. A good correlation has been observed between neutron depth dose distribution in a water phantom obtained using MOSFETs in integral mode and a tissue equivalent (T.E.) ionisation chamber. Such MOSFET dosimeters are a promising tool for micro-macro dosimetry in Boron Neutron Capture Therapy (BNCT) and Fast Neutron Therapy (FNT). Paired MOSFETs with one of the dosimeters covered by B-10 have been applied for measuring of average boron dose distribution and microdosimetric spectra due to alpha particles and Li-7 ions throughout a perspex phantom exposed in the epithermal neutron beam at the Brookhaven Medical Research Reactor (BMRR).
引用
收藏
页码:2693 / 2700
页数:8
相关论文
共 21 条
[1]   PMOS DOSIMETERS - LONG-TERM ANNEALING AND NEUTRON RESPONSE [J].
BLAMIRES, NG ;
TOTTERDELL, DHJ ;
HOLMES-SIEDLE, AG ;
ADAMS, L .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (06) :1310-1315
[2]  
BRABY LA, 1984, RADIAT PROT DOSIM, P219
[3]   A new radiotherapy surface dose detector: The MOSFET [J].
Butson, MJ ;
Rozenfeld, A ;
Mathur, JN ;
Carolan, M ;
Wong, TPY ;
Metcalfe, PE .
MEDICAL PHYSICS, 1996, 23 (05) :655-658
[4]   Device simulation of charge collection and single-event upset [J].
Dodd, PE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (02) :561-575
[5]  
LIU HB, 1996, IN PRESS NUCLEAR TEC
[6]   A SUPERCONDUCTING CYCLOTRON FOR NEUTRON RADIATION-THERAPY [J].
MAUGHAN, RL ;
POWERS, WE ;
BLOSSER, HG .
MEDICAL PHYSICS, 1994, 21 (06) :779-785
[7]  
MAUGHAN RL, 1992, PHYS MED BIOL, V37, P1957
[8]   Single-event effects experienced by astronauts and microelectronic circuits flown in space [J].
McNulty, PJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (02) :475-482
[9]  
NEMETS OF, 1975, HDB NUCLEAR PHYSICS, P101
[10]  
OLDHAM TR, IEEE T NUC SCI, V40, P1820