Efficient compensation of low-frequency magnetic field disturbances in NMR with fluxgate sensors

被引:18
作者
Kazimierczuk, K [1 ]
Kozminski, W [1 ]
机构
[1] Warsaw Univ, Dept Chem, PL-02093 Warsaw, Poland
关键词
compensation of magnetic field disturbances; fluxgate sensors; field-frequency lock; correlation spectroscopy; H-2; NMR;
D O I
10.1016/j.jmr.2005.02.018
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple stabilization scheme of B-0 magnetic field fluctuations is described. The method is based on external measurements of time dependent magnetic field fluctuations by fluxgate sensors and generation of a compensating correction current in a coil mounted directly on an NMR magnet. It is shown that such an approach efficiently eliminates relatively slow magnetic field variations with frequency up to approximately 100 Hz. In combination with a standard 2 H field-frequency lock system, the method enables acquisition of reproducible lineshapes and dramatically improves overall performance of a high resolution NMR spectrometer. The presented solution might substitute for the internal lock system in these case where deuterium lock is not available. (c) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:287 / 291
页数:5
相关论文
共 6 条
[1]   Analyzing and correcting spectrometer temperature sensitivity [J].
Bowyer, PJ ;
Swanson, AG ;
Morris, GA .
JOURNAL OF MAGNETIC RESONANCE, 2001, 152 (02) :234-246
[2]   Monitoring NMR spectrometer performance during data accumulation for macromolecular structure determination [J].
Braun, D ;
Wider, G ;
Wuthrich, K .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 110 (03) :313-315
[3]   SOURCES OF T1 NOISE IN TWO-DIMENSIONAL NMR [J].
MEHLKOPF, AF ;
KORBEE, D ;
TIGGELMAN, TA ;
FREEMAN, R .
JOURNAL OF MAGNETIC RESONANCE, 1984, 58 (02) :315-323
[4]   SYSTEMATIC SOURCES OF SIGNAL IRREPRODUCIBILITY AND T(1) NOISE IN HIGH-FIELD NMR SPECTROMETERS [J].
MORRIS, GA .
JOURNAL OF MAGNETIC RESONANCE, 1992, 100 (02) :316-328
[5]   REVIEW OF FLUXGATE SENSORS [J].
RIPKA, P .
SENSORS AND ACTUATORS A-PHYSICAL, 1992, 33 (03) :129-141
[6]   Advances in fluxgate sensors [J].
Ripka, P .