Potentiometric sensor for on line glucose determination

被引:8
作者
Folly, ROM
Valdman, B
Valero, F
Sola, C
机构
[1] UNIV FED RIO DE JANEIRO,ESCOLA QUIM,DEPT ENGN QUIM,RIO JANEIRO,BRAZIL
[2] UNIV AUTONOMA BARCELONA,DEPT ENGN QUIM,BELLATERRA 08193,SPAIN
关键词
D O I
10.1007/BF00154674
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An enzymatic biosensor, with a flow-through sensor for continuous monitoring of glucose coupled to a Flow Injection Analysis (FIA) sampling line has been used to determine glucose over the range of 5 to 120 g/l. without sample dilution. The system can analyse 10 samples per hour. Calibration and reproducibility analysis have shown a good linearity and excellent results as compared to a commercial glucose analyser.
引用
收藏
页码:867 / 870
页数:4
相关论文
共 5 条
[1]   LOW-LEVEL ELECTROCHEMICAL DETECTION OF GLUCOSE-OXIDASE AND A GLUCOSE-OXIDASE CONJUGATE [J].
ALLEN, RM ;
BENNETTO, HP ;
LEWIS, IAS ;
MASON, JR ;
WESTRA, AR .
BIOSENSORS & BIOELECTRONICS, 1995, 10 (6-7) :621-631
[3]  
Ruzicka Jaromir., 1988, FLOW INJECTION ANAL, V2nd
[4]  
VALDMAN B, 1993, ENZITEC 1 SEM NAC TE, P44
[5]   ONLINE FERMENTATION MONITORING USING FLOW-INJECTION ANALYSIS [J].
VALERO, F ;
LAFUENTE, J ;
POCH, M ;
SOLA, C ;
ARAUJO, AN ;
LIMA, JLFC .
BIOTECHNOLOGY AND BIOENGINEERING, 1990, 36 (06) :647-651