Transurethral prostate ablation using saline-liquid electrode introduced via flexible cystoscope

被引:11
作者
Hoey, MF
Dixon, CM
Paul, S
机构
[1] Univ Minnesota, Sch Med, Dept Physiol, Minneapolis, MN 55455 USA
[2] Univ Minnesota, Dept Urol, Minneapolis, MN 55455 USA
[3] Univ Minnesota, Dept Biomed Engn, Minneapolis, MN 55455 USA
[4] NYU, Dept Urol, New York, NY USA
关键词
D O I
10.1089/end.1998.12.461
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
An interstitial saline-electrode has been developed to couple radiofrequency (RF) energy to prostate tissue in order to produce large lesions quickly and controllably, In this study, we attempted to produce similar results using a flexible catheter introduced transurethrally via a flexible cystoscope. An insulated 4F catheter with an extendable needle electrode (26-gauge, 5-mm exposed metal tip) was introduced through a port of a flexible cystoscope into the urethra of 10 dogs via a small perineal incision (required for the canine U-shaped urethra). Under visual endoscopic guidance, the electrode was placed in the center of the prostate and the needle electrode extended at an angle into the prostate, one lobe at a time. Hypertonic saline (14.6% NaCl) was infused at 2 mL/min through the needle electrode into the tissue for 30 seconds before and during RF application. The energy (50 W, 475 kHz) was delivered for 30, 45, 60, or 90 seconds (RFT System; U.S. Surgical Corporation). Prostate and urethral temperatures were monitored using thermocouples mounted on the catheter, There was an automatic high-temperature shut-off if the urethra reached >42 degrees C or if impedance reached >300 Omega during RF energy application. Prostate sizes ranged from 1.4 x 2.0 x 1.4 cm to 5.0 x 4.5 x 4.2 cm, In this group, there were no automatic shut-offs at any of the energy durations, and the lesions produced ranged from 1.3 x 1.3 x 1.1 cm (0.99 cm(3)) to 1.6 x 2.7 x 2.2 cm (5.03 cm(3)). The average ablation volumes produced were 1.76 cm(3) at 30 seconds' RF application, 2.42 cm(3) at 45 seconds, 3.96 cm(3) at 60 seconds, and 5.03 cm(3) at 90 seconds. Histologic examination revealed typical coagulation necrosis in the treatment zone without tissue desiccation. Introduction of the saline-electrode through a flexible cystoscope provides a minimally invasive procedure that quickly and controllably produces large lesions that may provide effective treatment for benign prostatic hyperplasia with less patient discomfort.
引用
收藏
页码:461 / 468
页数:8
相关论文
共 39 条
[1]   THERMODYNAMIC TECHNIQUE FOR THE QUANTIFICATION OF REGIONAL BLOOD-FLOW [J].
ADAMS, T ;
HEISEY, SR ;
SMITH, MC ;
STEINMETZ, MA ;
HARTMAN, JC ;
FRY, HK .
AMERICAN JOURNAL OF PHYSIOLOGY, 1980, 238 (05) :H682-H696
[2]  
BARRY MJ, 1990, UROL CLIN N AM, V17, P495
[3]  
BELOT G, 1993, EUR UROL, V23, P326
[4]  
BHANOT SM, 1995, UROLOGY, V45, P427, DOI 10.1016/S0090-4295(99)80011-9
[5]   TRANSURETHRAL MICROWAVE THERMAL THERAPY - PATHOLOGICAL FINDINGS IN THE CANINE PROSTATE [J].
BOSTWICK, DG ;
LARSON, TR .
PROSTATE, 1995, 26 (03) :116-122
[6]   3-YEAR FOLLOW-UP OF URINARY SYMPTOMS AFTER TRANSURETHRAL RESECTION OF THE PROSTATE [J].
BRUSKEWITZ, RC ;
LARSEN, EH ;
MADSEN, PO ;
DORFLINGER, T .
JOURNAL OF UROLOGY, 1986, 136 (03) :613-615
[7]  
DAHLSTRAND C, 1993, EUR UROL, V23, P292
[8]   CONTROLLED DESTRUCTION AND TEMPERATURE DISTRIBUTIONS IN BIOLOGICAL TISSUES SUBJECTED TO MONOACTIVE ELECTROCOAGULATION [J].
EREZ, A ;
SHITZER, A .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1980, 102 (01) :42-49
[9]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[10]   PROSTATIC TISSUE DESTRUCTION BY HIGH-INTENSITY FOCUSED ULTRASOUND - EXPERIMENTATION ON CANINE PROSTATE [J].
GELET, A ;
CHAPELON, JY ;
MARGONARI, J ;
THEILLERE, Y ;
GORRY, F ;
CATHIGNOL, D ;
BLANC, E .
JOURNAL OF ENDOUROLOGY, 1993, 7 (03) :249-253