Endoscopic use of fibrin adhesives - Problems when injecting through long catheters

被引:8
作者
Nagelschmidt, M [1 ]
机构
[1] Univ Cologne, Dept Surg 2, Biochem & Expt Div, D-51109 Cologne, Germany
来源
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES | 1999年 / 13卷 / 01期
关键词
catheter; delivery force; endoscopic administration; fibrin sealant;
D O I
10.1007/s004649900905
中图分类号
R61 [外科手术学];
学科分类号
摘要
When used endoscopically fibrin adhesives are injected through twin-lumen catheters, which can reach up to 180 cm in length. Because fibrinogen solutions have a high viscosity, significant delivery forces are necessary, resulting in discomfort for the operator. Therefore, the two predominant fibrin sealants were characterized with respect to their viscosity and the force needed for their injection. Viscosity was determined at 18 degrees C, 25 degrees C, and 37 degrees C in a micro-Ostwald viscosimeter. Additionally, the maximum forces needed for injection through a 27-cm and a 160-cm catheter were determined at 25 degrees C in an Instron materials testing machine. Compared with preparation A the viscosity of preparation B was 8.0-34.5 times higher at 18 degrees C, 4.6-13.8 times higher at 25 degrees C and 3.1-6.4 times higher at 37 degrees C. In consequence, the delivery forces were 1.5-2.5 times (27 cm probe) and 3.4-4.5 times (160 cm probe) the values determined for preparation A. For preparation B a maximum load of 3.8 kg was necessary for injection. Assuming that different adhesive preparations have the same effect, a preparation of lower viscosity seems to be more suitable for use via long catheters than a preparation of high viscosity.
引用
收藏
页码:80 / 82
页数:3
相关论文
共 9 条
[1]   Prospective, randomized evaluation of the efficacy of fibrin sealant as a topical hemostatic agent at the cannulation site in neonates undergoing extracorporeal membrane oxygenation [J].
Atkinson, JB ;
Gomperts, ED ;
Kang, R ;
Lee, M ;
Arensman, RM ;
Bartlett, RH ;
RaisBharami, K ;
Breaux, CW ;
Cornish, JD ;
Haase, GM ;
Roden, J ;
Zwischenberger, JB .
AMERICAN JOURNAL OF SURGERY, 1997, 173 (06) :479-484
[2]   COMPARATIVE-STUDY OF BIOLOGICAL GLUES - CRYOPRECIPITATE GLUE, 2-COMPONENT FIBRIN SEALANT, AND FRENCH GLUE [J].
BASU, S ;
MARINI, CP ;
BAUMANN, FG ;
SHIRAZIAN, D ;
DAMIANI, P ;
ROBERTAZZI, R ;
JACOBOWITZ, IJ ;
ACINAPURA, A ;
CUNNINGHAM, JN .
ANNALS OF THORACIC SURGERY, 1995, 60 (05) :1255-1262
[3]  
FRIEDRICHS O, 1996, FIBRINKLEBUNG ENDOSK
[4]  
FUCHS KH, 1994, TECHNIK FIBRINKLEBUN, P113
[5]  
GRUND KE, 1994, TECHNIK FIBRINKLEBUN, P133
[6]   Hemostatic efficacy of a fibrin sealant-based topical agent in a femoral artery injury model: A randomized, blinded, placebo-controlled study [J].
Jackson, MR ;
Friedman, SA ;
Carter, AJ ;
Bayer, V ;
Burge, JR ;
MacPhee, MJ ;
Drohan, WN ;
Alving, BM .
JOURNAL OF VASCULAR SURGERY, 1997, 26 (02) :274-280
[7]  
MANEGOLD BC, 1994, TECHNIK FIBRINKLEBUN
[8]   Randomised trial of single and repeated fibrin glue compared with injection of polidocanol in treatment of bleeding peptic ulcer [J].
Rutgeerts, P ;
Rauws, E ;
Wara, P ;
Swain, P ;
Hoos, A ;
Solleder, E ;
Halttunen, J ;
Dobrilla, G ;
Richter, G ;
Prassler, R .
LANCET, 1997, 350 (9079) :692-696
[9]  
SALM R, 1994, TECHNIK FIBRINKLEBUN, P121