A Monte Carlo study of the quality dependence factors of common TLD materials in photon and electron beams

被引:46
作者
Mobit, PN
Nahum, AE
Mayles, P
机构
[1] Inst Canc Res, Joint Dept Phys, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden NHS Trust, Sutton SM2 5PT, Surrey, England
关键词
D O I
10.1088/0031-9155/43/8/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A Monte Carlo simulation of the quality dependence of different TL materials, in the form of discs 3.61 mm in diameter and 0.9 mm thick, in radiotherapy photon beams relative to Co-60 gamma-rays has been performed. The beam qualities ranged from 50 kV to 25 MV x-rays. The TL materials were: CaF2, CaSO4, LiF and Li2B4O7 The effects of the dopants on energy deposition in the TL material have also been determined for the highly sensitive LiF:Mg:Cu:P (TLD-100H) and for CaF2:Mn. It was found that there was a significant difference in the quality dependence factor derived from Monte Carlo simulations between LiF and LiF:Mg:Cu:P but not between CaF2 and CaF2:Mn. The quality dependence factors for Li2B4O7 varied from 0.990 +/- 0.008 (1 sd) for 25 MV x-rays to 0.940 +/- 0.009 (1 sd) for 50 kV x-rays relative to Co-60 gamma-rays; Monte Carlo simulations were also performed for Li2B4O7 in megavoltage electron beams. For CaF2, the quality dependence factor varied from 0.927 +/- 0.008 (1 sd) for 25 MV x-rays to 10.561 +/- 0.008 (1 sd) for 50 kV x-rays. The figure for CaSO4 ranged from 0.943 +/- 0.008 (1 sd) for 25 MV x-rays to 9.010 +/- 0.008 (1 sd) for 50 kV x-rays. The quality dependence factor for CaF2 increases by up to 5% with depth and by up to 15% with field size for the kilovoltage x-ray beams. For LiF-TLD, however, there was no significant dependence on the field size or depth of irradiation in the kilovoltage energy range.
引用
收藏
页码:2015 / 2032
页数:18
相关论文
共 40 条
[1]  
[Anonymous], P IAEA INT S MEAS AS
[2]  
AZORIN J, 1990, RADIAT PROT DOSIM, V33, P283
[3]   CORRECTION FACTORS FOR THICK-WALLED IONIZATION CHAMBERS IN POINT-SOURCE PHOTON BEAMS [J].
BIELAJEW, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (04) :501-516
[4]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[5]  
*BJR, 1983, BR J RADIOL S, V17
[6]   THE DEPENDENCE OF MASS-ENERGY ABSORPTION-COEFFICIENT RATIOS ON BEAM SIZE AND DEPTH IN A PHANTOM [J].
CUNNINGHAM, JR ;
WOO, M ;
ROGERS, DWO ;
BIELAJEW, AF .
MEDICAL PHYSICS, 1986, 13 (04) :496-502
[7]  
DEPLANQUE G, 1986, RADIAT PROT DOSIM, V17, P193
[8]  
DUANE S, 1989, PIRS0173 NAT RES COU
[9]  
EHRLICH M, 1980, 1119 NBS