Quantitative injection from a microloop. Reproducible volumetric sample introduction in capillary zone electrophoresis

被引:20
作者
Dasgupta, PK [1 ]
Surowiec, K [1 ]
机构
[1] MARIE CURIE SKLODOWSKA UNIV,DEPT CHEM PHYS,LUBLIN,POLAND
关键词
D O I
10.1021/ac951123i
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A wire loop deployed at the tip of a capillary electrophoresis system has been investigated as a means of quantitative injection. A thin film of a liquid is formed on the loop, the loop is transferred to a sealed chamber, and then pneumatic pressure is applied to introduce the contents of the loop into the capillary. As long as the applied pressure is below a certain threshold, no air is introduced into the capillary, even after the loop contents have been fully introduced. Sample surface tension and viscosity do not have a significant effect on the injected volume. The small loop injection technique appears to be a robust and reproducible alternative to presently practiced approaches to sample injection in CE.
引用
收藏
页码:1164 / 1168
页数:5
相关论文
共 16 条
[1]  
ADAMSON AW, 1976, PHYSICAL CHEM SURFAC, P21
[2]   MEASUREMENT OF GASES BY A SUPPRESSED CONDUCTOMETRIC CAPILLARY ELECTROPHORESIS SEPARATION SYSTEM [J].
DASGUPTA, PK ;
KAR, S .
ANALYTICAL CHEMISTRY, 1995, 67 (21) :3853-3860
[3]   ELECTRIC SAMPLE SPLITTER FOR CAPILLARY ZONE ELECTROPHORESIS [J].
DEML, M ;
FORET, F ;
BOCEK, P .
JOURNAL OF CHROMATOGRAPHY, 1985, 320 (01) :159-165
[4]   SPONTANEOUS INJECTION IN MICROCOLUMN SEPARATIONS [J].
FISHMAN, HA ;
AMUDI, NM ;
LEE, TT ;
SCHELLER, RH ;
ZARE, RN .
ANALYTICAL CHEMISTRY, 1994, 66 (14) :2318-2329
[5]  
FORET F, 1993, CAPILLARY ZONE ELECT, P140
[6]  
*ISC INC, OP MAN MODEL 3850 EL
[7]   ZONE ELECTROPHORESIS IN OPEN-TUBULAR GLASS-CAPILLARIES [J].
JORGENSON, JW ;
LUKACS, KD .
ANALYTICAL CHEMISTRY, 1981, 53 (08) :1298-1302
[8]   DURABLE GOLD-COATED FUSED-SILICA CAPILLARIES FOR USE IN ELECTROSPRAY MASS-SPECTROMETRY [J].
KRIGER, MS ;
COOK, KD ;
RAMSEY, RS .
ANALYTICAL CHEMISTRY, 1995, 67 (02) :385-389
[9]   ELECTROOSMOTICALLY PUMPED CAPILLARY FLOW-INJECTION ANALYSIS - VALVE-BASED INJECTION SYSTEMS AND SAMPLE THROUGHPUT [J].
LIU, SR ;
DASGUPTA, PK .
ANALYTICA CHIMICA ACTA, 1993, 283 (02) :739-745