Development of phantoms for spiral CT

被引:7
作者
Goodenough, DJ [1 ]
Levy, JR [1 ]
Kasales, C [1 ]
机构
[1] George Washington Univ, Dept Radiol, Washington, DC 20037 USA
关键词
phantoms; CT; image quality; spiral CT;
D O I
10.1016/S0895-6111(98)00020-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Purpose: This paper reports on the development of a new phantom for spiral CT. The phantom meets the increased demands on phantom z-axis uniformity in order that objects from the CT slice, immediately above and below the CT slice of interest, do not contribute perturbing information to the reconstructed CT slice. Material and methods: The phantom depends on formulation of tissue-like materials that can be cast and produced in both geometric and anthropomorphic shapes with sufficient z-axis length to enable unperturbed CT slices of test objects of interest. These materials are then used to produce a series of test objects of CT image quality including low contrast samples that do not require volume averaging or mixing of solutions, and that can reflect sub-slice thickness lest objects and supra-slice thickness test objects. Results: The overall phantom and its individual test objects provides meaningful tests of spiral CT image quality including slice sensitivity, CT number linearity and tests of high and low contrast resolution. Schematic designs and actual CT scans are shown. Conclusion: The new spiral phantom appears to meet the increased demands of spiral CT on phantom design, particularly z-axis length, and requirements for low contrast resolution test objects. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:247 / 255
页数:9
相关论文
共 14 条
[1]
COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION [J].
CRAWFORD, CR ;
KING, KF .
MEDICAL PHYSICS, 1990, 17 (06) :967-982
[2]
AN MTF METHOD IMMUNE TO ALIASING [J].
DROEGE, RT ;
RZESZOTARSKI, MS .
MEDICAL PHYSICS, 1985, 12 (06) :721-725
[3]
GOODENOUGH D, 1981, AM J NEURORADIOL, V2, P585
[4]
FACTORS RELATED TO LOW CONTRAST RESOLUTION IN CT SCANNERS [J].
GOODENOUGH, DJ ;
WEAVER, KE .
COMPUTERIZED RADIOLOGY, 1984, 8 (05) :297-308
[5]
DEVELOPMENT OF A PHANTOM FOR EVALUATION AND ASSURANCE OF IMAGE QUALITY IN CT SCANNING [J].
GOODENOUGH, DJ ;
WEAVER, KE ;
DAVIS, DO .
OPTICAL ENGINEERING, 1977, 16 (01) :52-65
[6]
A NEW SOFTWARE CORRECTION APPROACH TO VOLUME AVERAGING ARTIFACTS IN CT [J].
GOODENOUGH, DJ ;
WEAVER, KE ;
COSTARIDOU, H ;
EERDMANS, H ;
HUYSMANS, P .
COMPUTERIZED RADIOLOGY, 1986, 10 (2-3) :87-98
[7]
POTENTIAL ARTIFACTS ASSOCIATED WITH SCANNING PATTERN OF EMI SCANNER [J].
GOODENOUGH, DJ ;
WEAVER, KE ;
DAVIS, DO .
RADIOLOGY, 1975, 117 (03) :615-619
[8]
GOODENOUGH DJ, 1981, TECHNICAL ASPECTS CO, V5, P3993
[9]
SPIRAL(HELICAL) CT [J].
HEIKEN, JP ;
BRINK, JA ;
VANNIER, MW .
RADIOLOGY, 1993, 189 (03) :647-656
[10]
KRIZ RJ, SPIE, V555, P195