Reproducibility of 3D free-breathing magnetic resonance coronary vessel wall imaging

被引:32
作者
Desai, MY
Lai, SH
Barmet, C
Weiss, RG
Stuber, M
机构
[1] Johns Hopkins Univ, Sch Med, Russell H Morgan Dept Radiol & Radiol Sci, Baltimore, MD 21287 USA
[2] Johns Hopkins Univ, Sch Med, Dept Med, Baltimore, MD 21287 USA
[3] Johns Hopkins Univ, Sch Med, Dept Pathol, Baltimore, MD 21287 USA
[4] Johns Hopkins Univ, Sch Med, Dept Elect Engn, Baltimore, MD 21287 USA
[5] NIH, Bethesda, MD 20892 USA
[6] Univ Zurich, Inst Biomed Engn, Zurich, Switzerland
[7] ETH, Zurich, Switzerland
关键词
coronary artery vessel wall imaging; magnetic resonance imaging; reproducibility;
D O I
10.1093/eurheartj/ehi357
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Aims Although the coronary artery vessel wall can be imaged non-invasively using magnetic resonance imaging (MRI), the in vivo reproducibility of wall thickness measures has not been previously investigated. Using a refined magnetization preparation scheme, we sought to assess the reproducibility of three-dimensional (3D) free-breathing black-blood coronary MRI in vivo. Methods and results MRI vessel wall scans parallel to the right coronary artery (RCA) were obtained in 18 healthy individuals (age range 25-43, six women), with no known history of coronary artery disease, using a 3D dual-inversion navigator-gated black-blood spiral imaging sequence. Vessel wall scans were repeated 1 month later in eight subjects. The visible vessel wall segment and the wall thickness were quantitatively assessed using a semi-automatic tool and the intra-observer, inter-observer, and inter-scan reproducibilities were determined. The average imaged length of the RCA vessel wall was 44.5 +/- 7 mm. The average wall thickness was 1.6 +/- 0.2 mm. There was a highly significant intra-observer (r=0.97), inter-observer (r=0.94), and inter-scan (r=0.90) correlation for wall thickness (all P < 0.001). There was also a significant agreement for intra-observer, inter-observer, and inter-scan measurements on Bland-Altman analysis. The intra-class correlation coefficients for intra-observer (r=0.97), inter-observer (r=0.92), and inter-scan (r=0.86) analyses were also excellent. Conclusion The use of black-blood free-breathing 3D MRI in conjunction with semi-automated analysis software allows for reproducible measurements of right coronary arterial vessel-wall thickness. This technique may be well-suited for non-invasive longitudinal studies of coronary atherosclerosis.
引用
收藏
页码:2320 / 2324
页数:5
相关论文
共 28 条
[1]
INTRACLASS CORRELATION COEFFICIENT AS A MEASURE OF RELIABILITY [J].
BARTKO, JJ .
PSYCHOLOGICAL REPORTS, 1966, 19 (01) :3-&
[2]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]
Improved coronary artery definition with T2-weighted, free-breathing, three-dimensional coronary MRA [J].
Botnar, RM ;
Stuber, M ;
Danias, PG ;
Kissinger, KV ;
Manning, WJ .
CIRCULATION, 1999, 99 (24) :3139-3148
[4]
Botnar RM, 2001, MAGNET RESON MED, V46, P848, DOI 10.1002/mrm.1268
[5]
Prospective navigator correction of image position for coronary MR angiography [J].
Danias, PG ;
McConnell, MV ;
Khasgiwala, VC ;
Chuang, ML ;
Edelman, RR ;
Manning, WJ .
RADIOLOGY, 1997, 203 (03) :733-736
[6]
Deyo RA, 1991, CONTROLLED CLIN TRIA, V12, P142
[7]
Soap-bubble visualization and quantitative analysis of 3D coronary magnetic resonance angiograms [J].
Etienne, A ;
Botnar, RM ;
van Muiswinkel, AMC ;
Boesiger, P ;
Manning, WJ ;
Stuber, M .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (04) :658-666
[8]
Fayad ZA, 2000, CIRCULATION, V102, P506
[9]
Fischer SE, 1999, MAGN RESON MED, V42, P361, DOI 10.1002/(SICI)1522-2594(199908)42:2<361::AID-MRM18>3.0.CO
[10]
2-9