Co-immobilized L-lactate oxidase and L-lactate dehydrogenase on a film mounted on oxygen electrode for highly sensitive L-lactate determination

被引:20
作者
Casimiri, V
Burstein, C
机构
关键词
L-lactate biosensor; L-lactate recycling; L-lactate oxidase; L-lactate dehydrogenase co-immobilization; E-coli L-lactate determination;
D O I
10.1016/0956-5663(96)85929-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Covalent immobilization of L-lactate oxidase (LOD) with L-lactate dehydrogenase (LDH) on a film tightly bound to an oxygen electrode, for rapid and sensitive L-lactate measurements, is described. Regeneration of L-lactate by substrate recycling provided an amplification of the sensor response, making it possible to decrease the detection limit of L-lactate from 10 mu M to 20 nM. The apparent K-m for L-lactate with the LOD-LDH coupled reaction was 1 mu M, compared with 3 mM when utilizing only LOD. Linearity was obtained from 20 to 300 nM with both enzymes, whereas with LOD alone it was from 10 mu M to 1 mM. Optimization of the biosensor was obtained with an increase in LOD and LDH film loading and low L-lactate concentration. The enzymes covalently bound to the him stabilized the biosensor (half life 8 weeks) for over 400 measurements. Low L-lactate excreted by E. coli bacteria metabolism can be assayed in turbid culture medium without pretreatment by the amplified L-lactate detection. (C) 1996 Elsevier Science Limited
引用
收藏
页码:783 / 789
页数:7
相关论文
共 16 条