Evidence for a significant timer error on a linear accelerator: consequences for a special therapy application
被引:11
作者:
Biggs, PJ
论文数: 0引用数: 0
h-index: 0
机构:
Massachusetts Gen Hosp, Harvard Med Sch, Dept Radiat Oncol, Div Radiat Biophys, Boston, MA 02114 USAMassachusetts Gen Hosp, Harvard Med Sch, Dept Radiat Oncol, Div Radiat Biophys, Boston, MA 02114 USA
Biggs, PJ
[1
]
机构:
[1] Massachusetts Gen Hosp, Harvard Med Sch, Dept Radiat Oncol, Div Radiat Biophys, Boston, MA 02114 USA
A significant timer error has been found on a linear accelerator used for intraoperative radiation therapy. Since typical treatment doses range up to 20 Gy, calibrations that do not account for this effect can incur a large dosimetry error. The effect is dose rate dependent, but energy independent. For a dose rate of 900 cGy min(-1), the error in dose calibration varies between 2 and 3% for an average of 2.5%. At dose rates of 600 and 300 cGy min(-1), the error was 1.7% and 0.8% respectively. The average timer error at 900 cGy min(-1) was found to be 1.2 +/- 0.1 monitor units. It is argued that even a small timer error has serious dosimetry consequences for the implementation of intensity modulated radiation therapy.