The purpose of this study was to evaluate a new device specifically designed for patent ductus arteriosus (PDA) occlusion based on PDA anatomy. Percutaneous closure of aortopulmonary shunts was attempted in 10 dogs. Shunts were surgically created in the location and orientation of PDA. The occlusion prosthesis consisted of a cylindrical frame filled with polyester and a 32 degrees angled retention disk. The delivery system of the device included a 6 Fr thin-walled Teflon introducing sheath, an antirotating delivery catheter, and cable. Complete occlusion of the shunt was obtained in nine animals. One animal died before device placement. Temporary hemolysis occurred in one dog after device placement. The disk of device was completely covered by smooth glistening neoendothelium at 1-3 months postmortem examination. None of the devices protruded into the lumen of the aorta. The device is an improvement over the standard Amplatzer duct occludes. The angled retention disk lies flatly against the aortic wall, avoiding turbulence or an aortic pressure gradient. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:571 / 576
页数:6
相关论文
共 35 条
[1]
AMPLATZ K, 1993, RADIOLOGY CONGENITAL, P291
[2]
ARGUELLES JIZ, ANAL ESPANOLES EDIAT, V44, P332
[3]
ARGUELLES JIZ, 1966, ANAL ESPANOLES PEDIA, V44, P332
[4]
BECKER AE, 1981, PATHOLOGY CONGENITAL, P349
[5]
CHISHOLM JC, 1995, PEDIATR CARDIOL, V16, P194
[6]
Echocardiographic and radionuclide pulmonary blood flow patterns after transcatheter closure of patent ductus arteriosus [J].
Dessy, H ;
Hermus, JPS ;
vandenHeuvel, F ;
Oei, HY ;
Krenning, EP ;
Hess, J .
CIRCULATION, 1996, 94 (02) :126-129
[7]
A PHYSIOLOGICAL AND ANATOMICAL STUDY OF THE CLOSURE OF THE DUCTUS ARTERIOSUS IN THE DOG [J].
EVERETT, NB ;
JOHNSON, RJ .
ANATOMICAL RECORD, 1951, 110 (01) :103-111
[8]
LEFT PULMONARY-ARTERY STENOSIS AFTER TRANSCATHETER OCCLUSION OF PERSISTENT ARTERIAL DUCT [J].
FADLEY, F ;
ALHALEES, Z ;
GALAL, O ;
KUMAR, N ;
WILSON, N .
LANCET, 1993, 341 (8844) :559-560
[9]
Retrievable patent ductus arteriosus plug for interventional, transvenous occlusion of the patent ductus arteriosus - Evaluation in lambs and preliminary clinical results [J].
Grabitz, RG ;
Schrader, R ;
Sigler, M ;
Seghaye, MC ;
Dzionsko, C ;
Handt, S ;
Schneidt, B ;
VonBernuth, G .
INVESTIGATIVE RADIOLOGY, 1997, 32 (09) :523-528
[10]
TRANSCATHETER OCCLUSION OF THE PERSISTENTLY PATENT DUCTUS-ARTERIOSUS - 40-MONTH FOLLOW-UP AND PREVALENCE OF RESIDUAL SHUNTING [J].
HOSKING, MCK ;
BENSON, LN ;
MUSEWE, N ;
DYCK, JD ;
FREEDOM, RM .
CIRCULATION, 1991, 84 (06) :2313-2317