Parallel magnetic resonance imaging

被引:17
作者
Katscher, Ulrich [1 ]
Boernert, Peter [1 ]
机构
[1] Philips Res Europe Hamburg Lab, D-22335 Hamburg, Germany
关键词
magnetic resonance imaging; parallel imaging; parallel transmission; RF coil array; RF pulses; inverse problem;
D O I
10.1016/j.nurt.2007.04.011
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Parallel MRI started with the introduction of coil arrays in improving radiofrequency (RF) acquisition (what is called parallel imaging) and continued with an analogous development for RF transmission (parallel transmission). Based on differences in the spatial sensitivity distributions of the involved array elements, both techniques try to shorten the respective k-space trajectory. Parallel imaging refers to the acquisition of k-space data, whereas parallel transmission is dealing with the deposition of RF energy packages in the excitation k-space. However, parallel transmission is not simply the reciprocal of parallel imaging. The main goal of parallel imaging is the shortening of the acquisition time. The main goal of parallel transmission is the shortening of the pulse duration of spatially selective RF pulses. The present article describes the basic concept, the state of the art, and the similarities and differences of both technologies.
引用
收藏
页码:499 / 510
页数:12
相关论文
共 117 条
[1]   Transmit and receive transmission line arrays for 7 tesla parallel imaging [J].
Adriany, G ;
Van de Moortele, PF ;
Wiesinger, F ;
Moeller, S ;
Strupp, JP ;
Andersen, P ;
Snyder, C ;
Zhang, XL ;
Chen, W ;
Pruessmann, KP ;
Boesiger, P ;
Vaughan, T ;
Ugurbil, K .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (02) :434-445
[2]  
ALEXANDER A, 2007, NEUROTHERAPEUTICS
[3]   Coil-based artifact reduction [J].
Atkinson, D ;
Larkman, DJ ;
Batchelor, PG ;
Hill, DLG ;
Hajnal, JV .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (04) :825-830
[4]  
Augustin Michael, 2004, Top Magn Reson Imaging, V15, P207, DOI 10.1097/01.rmr.0000132790.61835.f7
[5]   Improved diffusion-weighted single-shot echo-planar imaging (EPI) in stroke using Sensitivity Encoding (SENSE) [J].
Bammer, R ;
Keeling, SL ;
Augustin, M ;
Pruessmann, KP ;
Wolf, R ;
Stollberger, R ;
Hartung, HP ;
Fazekas, F .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (03) :548-554
[6]   A simple method for investigating the effects of non-uniformity of radiofrequency transmission and radiofrequency reception in MRI [J].
Barker, GJ ;
Simmons, A ;
Arridge, SR ;
Tofts, PS .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (841) :59-67
[7]  
Börnert P, 1998, MAGN RESON MATER PHY, V7, P166
[8]  
BOLLENBECK J, 2005, P INT SOC MAGN RESON, V13, P860
[9]   Curved slice imaging [J].
Bornert, P ;
Schaffter, T .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (06) :932-939
[10]  
BOSKAMP E, 2002, P INT SOC MAGN RESON, V10, P903